148
FLIM Data Acquisition
Figure 134:
Selecting the white light laser pulse frequency in Beam Path Settings
If the repetition rate is changed, it can become necessary to adapt the resolution in
SymPhoTime (see
16.2.7
Setting the Pinhole
On the
Acquisition
tab in LAS AF, set the pinhole diameter (
Pinhole
) to
1 Airy
for FLIM
measurements with VIS and UV excitation and to the maximum value for MP excitation.
16.2.8
Optimizing FLIM Settings
1. First, find the position of your specimen you want to use for FLIM measurement.
2. Click on the image acquisition button
Run FLIM Test
to start the scanning operation. You
can view and optimize the fluorescence intensity in the
Count Rate Monitor
on the
Setup
tab (
3. Laser line and intensity can be modified during the test measurement until you attain the
count rate desired.
4. When you start the test measurement, an online FLIM image is displayed in
SymPhoTime.
Do not select other frequencies on the laser
If you select
manual
in the pull-down menu, you can select additional
frequencies on the laser yourself. This is, however, not recommended
because the cable lengths are not calibrated for these frequencies.
Scan speeds 100 to 200 Hz recommended
Do not use scan speeds that are too high because these can lead to
longer analysis times in SymPhoTime. Scan speeds at 100 to 200 Hz are
recommended. Avoid the use of the resonant scanner for FLIM data
acquisition as much as possible.
The online FLIM image is not saved.
This image is not saved. To save your data, you have to work in the
Measurements
step.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......