141
FLIM Data Acquisition
Figure 124:
Beam Path Settings in LAS AF: Activating the continuous wave VIS lasers
If the system has an FCS filter wheel, its setting is irrelevant for FLIM, as it is directly in front
of the argon laser and thus does not affect the pulsed excitation.
16.2.2.2
Using Pulsed Diode Lasers (UV, VIS)
• For external FLIM with APDs or an external photomultiplier: Follow the instructions in
. The fluorifier disc can be set to
Empty
as needed. If additional suppression
of the excitation light is desired, follow the recommended settings in
• For internal SP FLIM with SP FLIM detectors: Check the setting of the fluorifier disc. It
contains an additional barrier filter that prevents excitation light from hitting the SP
detection module. Recommended settings are given in
in
16.2.2.3
Using MP Lasers
• For external FLIM at the X1 port and for internal FLIM: Follow the instructions in
. With the fluorifier disc, an MP filter block has to be selected (SP680 or
SP700); the pinhole should be open as far as possible.
• For FLIM detection at NDD position (HyD RLD): The setting of the fluorifier disc, X1 port,
and
pinhole
play no role.
Prevent light incidence during MP FLIM measurement
During an MP FLIM measurement (particularly with HyD RLD), the light in
the room should be switched off and the shutter of the fluorescence lamp
should be closed. Furthermore, the iris diaphragm on the condenser
should be completely closed.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......