Instrument
error
Effects
Hz
Effects
V
Elimination
with two face
measurement
Automatically
corrected
with proper
adjustment
c - Line of sight
error
ü
---
ü
ü
a - Tilting axis
error
ü
---
ü
ü
l - Compensator
index error
---
ü
ü
ü
t - Compensator
index error
ü
---
ü
ü
i - Vertical index
error
---
ü
ü
ü
5.2
Preparation
Before determining the instrument errors, the instrument has to
be levelled using the electronic level.
The tribrach, the tripod and the underground should be stable
and secure from vibrations or other disturbances.
The instrument should be protected from direct sunlight to
avoid thermal warming.
It is also recommended to avoid strong heat shimmer and air
turbulence. The best conditions are early in the morning or with
overcast sky.
Before starting to work, the instrument has to become acclimatised to the
ambient temperature. Approximately two minutes per °C of temperature differ-
ence from storage to working environment, but at least 15 min, should be
taken into account.
5.3
Combined Adjustment (l, t, i and c)
The combined adjustment procedure determines the following instrument
errors in one process:
Instrument error
Description
l, t
Compensator longitudinal and transversal index
errors
i
Vertical index error, related to the standing axis
c
Horizontal collimation error, also called line of
sight error
The following table explains the most common settings.
1.
Leica Captivate - Home
:
Settings
\
TS instrument
\
Check & adjust
2.
Check & Adjust
Summary of errors to
be adjusted
electronically
☞
☞
☞
Description
Combined adjustment
procedure step-by-
step
Check & Adjust
39