FLEXACAM C1 User Manual
Stand-Alone Operation Installation / Setup / Application 28
General notes
The white balance ensures that the
specimen is shown in neutral colors. We
recommend carrying out a new white balance
adjustment whenever you change the illumina-
tion type or color temperature. This is particu-
larly necessary if you are working with halogen
light that can shift from yellow (low intensity)
to blue (high intensity).
Configure the white balance via the user
menu. Depending on the configuration,
automatic white balance can also be carried out
using the footswitch.
The white balance of the camera is
factory set to automatically provide
good initial results when used in conjunction
with Leica LED illumination.
Manual white balance
O
Manual white balance is described on
page 39.
One Push white balance
1. Position a gray chart or a gray, neutral
object under the microscope so that the
entire image field is filled.
2. Adjust the illumination as desired.
3. Press the "WB" button on OSD.
White Balance
Содержание FLEXACAM C1
Страница 1: ...FLEXACAM C1 12 MP Stand Alone Microscope Camera User Manual ...
Страница 4: ...FLEXACAM C1 User Manual General Notes 4 General Notes ...
Страница 12: ...FLEXACAM C1 User Manual Introduction 12 Introduction ...
Страница 17: ...FLEXACAM C1 User Manual Assembly 17 Assembly ...
Страница 45: ...FLEXACAM C1 User Manual Care and Maintenance 45 Care and Maintenance ...
Страница 50: ...FLEXACAM C1 User Manual Specifications 50 Specifications ...
Страница 52: ...FLEXACAM C1 User Manual Specifications 52 Dimensional Drawings Dimensions in mm ...
Страница 53: ...FLEXACAM C1 User Manual Troubleshooting 53 Troubleshooting ...