16
Microflash
The microflash is assembled in the same way
(only in conjunction with the switchable mirror
and a lamphousing).
Ventilation
Attention:
Imp ortant: Make sure that the instrument has
sufficient ventilation:
Take care not to block the air supply
underneath the microscope and at the con-
nected lamphousings or the air vents on the
top of the microscope with paper, etc. Fire
hazard! Minimum distance from inflammable
objects 10 cm (4˝).
Lamphousing 106, 106 z
Attention:
Only lamhousing 106 (48.1) can be used for
transmitted light!
Remove the dust protection cover from the lamp
mount. Unscrew the clamp screw (3.2) with the
aid of the hexagonal screwdriver (1.1) so that the
screw on the inner surface of the lamp mount
does not protrude above the surface. Align the
lamphousing so that the screw engages in the
corresponding indentation on the lamphousing.
Tighten the screw to fix the lamphousing firmly
to the microscope.
Filter mount
A filter mount (Fig. 9) taking up to four extra
filters (50 mm diameter) can be assembled
between the microscope and the lamphousing
in the same way. When lamphousing 106 is used,
only 1 thick or 2 thin filters can be inserted.
Fig. 8
Lamphousing 106 z with Hg 50 burner
1
Lid,
2
Collector,
3
Burner (Hg 50),
4
Reflector,
5, 7
x/y
adjustment of the reflector,
6
Reflector focusing,
8
Sockets for
safety cut-out plug,
9
Lamp holder screws
1
2
3
4
5
6
7
8
9
9
!