Leica CM1860/CM1860 UV
Safety
2
2.3.2 Safety guard
Warning
Prior to manipulating the knife and specimen, changing the specimen or taking a break, always lock the
handwheel and cover the cutting edge with the safety guard.
The knife holder CN and the premium blade holder feature safety guards; the glass anti-roll plate of the
CE and CE-TC blade holders also serves as a safety guard.
2.4 Operating Conditions
Transport and installation
•
After transporting, do not turn the instrument on for a minimum of 4 hours.
•
A waiting period is necessary to allow the compressor oil, which may have been displaced during
transport, to return to its original position. Failure to comply with this may cause damage to the
instrument.
•
The instrument may not be operated in hazardous locations!
•
To ensure proper functioning of the instrument, it must be set up while maintaining a minimum
distance on all sides from walls and furniture:
-
rear:
15
cm
-
right:
30
cm
-
left:
15
cm.
Electrical connection
•
Before connecting the instrument to an AC power outlet, ensure that the AC power of your laboratory
matches the values on the nameplate of the instrument!
•
The minimum rated voltage must be maintained during the startup phase of the cooling system -
! The compressor draws a starting current between 45 and 50 A. Therefore,
the electric circuit at the installation site must be inspected by an electrician to ensure that it meets
the requirements for a smooth operation of the instrument. A stable power supply in accordance with
the instrument's specifications is essential to its proper functioning. Failure to comply with the above
can damage the instrument.
•
No additional extension cord may be used.
Defrosting
Warning
The quick freeze shelf can become very hot during the defrosting process! Therefore, do not touch it!