18
Version 2.4, Revision L
Operation
5
5.
Operation
5.1
Installation site requirements
The place of installation must meet the following requirements:
• The instrument is designed for indoor use only.
• The power plug must be freely and easily accessible.
• Power supply at a distance no greater than the length of the power cable (3m) – an extension cable
must not be used.
• Level installation location,
• Substrate as free of vibration as possible,
• Relative humidity should not exceed 80 %
• Room temperature consistently b5 °C and +40 °C
• Avoid vibrations, direct sunlight, and large temperature fluctuations!
Warning
The instrument may not be operated in hazardous locations.
Warning
The instrument MUST be connected to a grounded power socket. Use only a provided power cable
that is intended for the local power supply.
5.2
Setting up the instrument
Note
The instrument MUST be connected to a grounded power socket. Use only a provided power cable
that is intended for the local power supply.
Warning
The instrument MUST be set up so that the power plug and switch are free and easily accessible at
all times!
Note
The VT1000 S is equipped with a autoranging power supply to cover voltages from 100 V to 240 V.
After switching on the main switch, the instrument carries out an initial startup run: The blade returns
to the rear starting position after a short forward movement.