Cleaning and maintenance
6
Checking the bottom tray
Caution
Some of the reagents used in the HistoCore PEARL are hazardous to health. Therefore always wear
chemical protective clothing, protective goggles, respirator, rubber gloves and all other necessary
personal protective equipment, when handling them.
The bottom tray must be inspected regularly for signs of leaking reagents. It is recommended to do this
after the processing program and cleaning program has been completed, before any leaking reagent
has a chance to evaporate.
To check the bottom tray, use one hand to pull the bottom tray out by the handle until you see the
etched number, 20 cm; use the other hand to support the tray from the bottom. Then pull out the whole
bottom tray.
Cleaning the paraffin drip tray
Caution
Use care with paraffin - risk of burning when removing the paraffin drip tray due to liquid paraffin.
•
Pull out the paraffin drip tray. Empty the paraffin drip tray with the supplied scraper.
•
If necessary place it in the laboratory oven (on cellulose) and clean with xylene substitute.
6.2.2 Weekly cleaning and maintenance
Cleaning the paraffin baths
Caution
Work carefully and wear gloves!
Be careful as the walls of the paraffin baths are very hot and may cause burns!
•
Wipe the paraffin baths and the lids using a lint-free cloth moistened with solvent (xylene or paraffin
remover).
•
The lid can be removed for cleaning purposes.
•
If they are very dirty, remove the paraffin filters from the paraffin baths. Clean, dry, and reinsert them.
•
The paraffin bath seals must be replaced yearly. Contact Leica Biosystems Technical Service for the
replacement service.
Emptying the condensate bottle
The condensate bottle must be taken out, emptied and reinserted weekly. This should be done sooner if
a prompt to that effect is issued by the software.
Caution
Dispose of waste solvents with care according to local regulations and the waste management policy of
the company or institution.