88
Version 1.3, Revision D
Cleaning and maintenance
6
Caution
Do not load unprocessed specimens or wet baskets into the retort prior to running a cleaning program.
Formalin and water residue can flow into the reagent bottle/paraffin bath from the last step of the previous
processing cycle.
If you have inadvertently loaded unprocessed specimens into the retort and have started a cleaning
program, stop the program, remove the unprocessed specimens and finish the cleaning program. Replace
all the processing reagents and paraffin before doing any further processing.
Caution
Baskets can be cleaned together in the retort. ONLY completely cleaned baskets can be used for tissue
processing. After a maximum of 5 cleaning cycles, cleaning xylene and cleaning ethanol must be replaced.
Neither drip trays nor other objects can be cleaned using the HistoCore PEARL Cleaning program!
•
After you touch the
OK
button, a message appears indicating that before the cleaning is started the
retort must be empty and all specimens have to be removed.
•
When all necessary conditions have been satisfied, press the
Start
button. Wait about 8 seconds, and
the cleaning program starts. The procedure is displayed graphically on the Dashboard.
•
After the cleaning program has been completed, a message appears on the screen if the temperature
in the retort is high. Open the retort lid and wait for a few minutes to let it cool down.
Cleaning reagent warning message
If RMS is enabled, the following indicates that the warning threshold values have been exceeded.
•
The exceeded bottle turns to red and flashes. The operator is notified that the reagent has exceeded
and can replace this reagent before the program starts. However, it is also possible to start the
program despite this at Supervisor level.
•
The exceeded bottle turns to red and flashes at the end of a program (after the cleaning). You can go
directly to the reagent
Status
window.
•
The reagent status highlights the exceeded values in red.
Caution
Follow the RMS reminder. Continuously reusing the expired cleaning reagents may lead to bad quality of
the Cleaning program and may cause damage to the instrument or trigger an error reporting.
6.2
General instrument cleaning
6.2.1 Daily cleaning and maintenance
Instrument exterior
Clean the housing (including the instrument exterior, display, retort lid, counter area, etc) and the
ground if contaminated by reagent. Wipe with a damp lint-free cloth moistened with mild detergent and
dry.