background image

IWT-5000A Handler I/O Control Mode

The handler I/O is for connecting external controller as automatic use. The external device can 
connect EXT_VCC (DC +5V) and EXT_GND. And via  ‘MODE’ , ’START’ signal to control the 
system test mode (by select sample/ test, and START). You can via handler I/O output PASS’ , 
‘FAIL’ , ‘BUSY’ signal to know IWT-5000A test result is PASS, FAIL. BUSY or finished. 

MODE’  &  ’START’ is the input signal. When the signal level too high over EXT_VCC 26%, the 

system will judge the signal is High. For example,  ‘EXT_VCC’=5V, ‘MODE’ > 1.3V, that is High, 
‘START’ < 1.2V, that is Low

PASS’ , ‘FAI;’ , ‘BUSY’  signals are Open collector output. You can via optical coupling 

component connect to single chip or PC I/O card for automatic control. Or directly connect to LED 
for display. Please accord the voltage and current to use the suitable resister for current-limit. 

You can refer the below external circuit diagram for do a external board. 

1.

Please use this diagram connect IWT-5000A handler I/O port (15Pin D SUB Connector)

2.

Connect the sample to IWT-5000A. Set the test voltage, DIV, AVG, etc. 

3.

Press [System] on the IWT-5000A, and press [F5] ’Ext I/O’. The system will enter external 
Handler I/O mode. 

4.

Switch the external board [SW1] to OFF. It's the sample mode. 

5.

Check if the external board BUSY LED is off. Press [SW2], do not release it, for testing. 
During the testing, the BUSY LED will light. After LED off, it means finishing the sampling. You 

Отзывы: