Specifications
OX-TRAN Model 2/22 Operator’s Manual
9-2
Revision R
MOCON, Inc.
Gas Supply Requirements
Carrier Gas
Gas Composition
Nitrogen/Hydrogen Mix (98%/2%)
Supply Pressure, Nominal
25 PSI, (1.72 Bar), (172 kPa)
Supply Pressure, Maximum
27 PSI, (1.86 Bar), (186 kPa)
Supply Pressure, Minimum
22 PSI, (1.52 Bar),
152 kPa)
Flow Rate, Nominal, Test Active
80 cc/minute
Test Gas
Gas Composition
100% Oxygen (purity 99.9%)
Supply Pressure, Nominal
25 PSI, (1.72 Bar), (172 kPa)
Supply Pressure, Maximum
29 PSI, (2 Bar), (200 kPa)
Supply Pressure, Minimum
20 PSI, (1.38 Bar), 138 kPa)
Flow Rate, Nominal, Test Active
35 cc/minute
Table 9.4: Gas Supply Specifications
Environmental Chamber Specifications
Temperature Range
5 °C to 50 °C ± 0.5 °
C
Humidity Control
None
Table 9.5: Environmental Chamber Specifications