background image

ENGLISH 

 

 

 

Fabrication Instructions

 

 

X-tension is designed to resist or stop dorsiflexion.  

 

 
                                                

Mounting holes                                      SNAPstop window

 

  

 

1)

 

If your orthosis requires a stop, sagittally align your AFO before installing X-tension. (SNAPstop is the only 
plantar flexion stop that fits the X-tension geometry) 

2)

 

Position the X-tension band on the posterior aspect of the AFO with the SNAPstop Window directly 
over the SNAPstop. (If SNAPstop is not used, position the bottom of the SNAPstop Window level with 
the axis of the ankle joint). 

3)

 

Mark the location of the two lower mounting holes on the plastic with a marking pencil or with hot 
poker.  

4)

 

Drill a 1/8” (3.2mm) hole at the marked points.  

5)

 

Secure the lower band with Speedy rivets (for Dynamic X-tension) or copper rivets (for Static X-
tension).  

6)

 

Once the two bottom holes are secured, stretch the X-tension band over the AFO and mark the two 
upper mounting holes. 

7)

 

Drill a 1/8” (3.2mm) hole at the marked points. To increase tension in the X-tension band, drill your 
holes slightly up and out from your marks.  

8)

 

Stretch the X-tension band slightly (if Dynamic) and align the holes. Secure the upper band with 
Speedy rivets (for Dynamic X-tension) or copper rivets (for Static X-tension). 
 

 

 

 

 

 

 

This product should only be used for its intended purpose and should be fabricated under  

the guidance of a certified Orthotist per the provided instructions. 

 

 