Installation and User Manual 

version 1.03 

 

TLB4 

 

 

2014/30/UE

 

 

EN55022:2010   EN61000-6-2:2005   EN61000-6-4:2007 

 

 

SYSTEM IDENTIFICATION

 

 

 

 

Содержание TLB4-ANA

Страница 1: ...ENGLISH ENGLISH ENGLISH ENGLISH Installation and User Manual version 1 03 TLB4 2014 30 UE EN55022 2010 EN61000 6 2 2005 EN61000 6 4 2007 SYSTEM IDENTIFICATION...

Страница 2: ...er No guarantee against misuse Batteries Laumas provides 1 year guarantee from the date of delivery note against material defects or battery manufacturing faults Disposal of Waste Equipment by Users i...

Страница 3: ...ERO SETTING 13 ZERO VALUE MANUAL ENTRY 14 REAL CALIBRATION WITH SAMPLE WEIGHTS 15 CONFIRMATION AND CHANGE OF ACTIVE CHANNELS 16 EQUALIZATION 17 FILTER ON THE WEIGHT 18 ANTI PEAK 18 ZERO PARAMETERS 19...

Страница 4: ...TIONAL MENU MAP 34 TLB4 REMOTE CONTROL 35 REMOTE KEYPAD LOCKING 35 ADDITIONAL ALARMS 35 ALARMS 36 PRINTING EXAMPLES 38 FRONT PANEL MOUNTING 40 RESERVED FOR THE INSTALLER 41 MENU LOCKING 41 MENU UNLOCK...

Страница 5: ...n solenoid valve and remote control switch coils Avoid inverters in the instrument panel if inevitable use special filters for the inverters and separate them with sheet metal partitions The panel ins...

Страница 6: ...TURE By means of a copper wire with suitable cross section connect the cell upper support plate with the lower support plate then connect all the lower plates to a single earthing system Electrostatic...

Страница 7: ...n wire and the negative excitation wire must be equal or similar to the one indicated in the load cell data sheet input resistance Load cell voltage measurement use a digital multimeter Turn on the in...

Страница 8: ...play printer Optional integrated Profibus DP DeviceNet CANopen Profinet IO Ethernet IP Ethernet TCP IP Modbus TCP RS232 RS485 SERCOS III PowerLink CC Link output TLB4 only optoisolated 16 bit analog o...

Страница 9: ...PER SECOND 600 conversions second DISPLAY RANGE 999999 NO OF DECIMALS DISPLAY INCREMENTS 0 4 x 1 x 2 x 5 x 10 x 20 x 50 x 100 DIGITAL FILTER READINGS PER SECOND 0 006 7 s 5 600 Hz RELAY LOGIC OUTPUTS...

Страница 10: ...of difficulty connecting all the reference wires of the installed load cells simply connect those of the load cell located at the average distance from the instrument The reference wires not used mus...

Страница 11: ...No 2 VDC min 5 V max 24 V 22 LOAD CELLS REF SENSE 7 INPUT COMMON VDC 0 V 23 LOAD CELL 3 SIGNAL 8 SUPPLY 12 24 VDC 24 LOAD CELL 3 SIGNAL 9 SUPPLY 12 24 VDC RS485 SHIELD GND 25 LOAD CELLS 3 and 4 EXCITA...

Страница 12: ...tion during weight display press and hold down the keys and press immediately followed by KEY Short press Long press 3 s Into menus Semi automatic zero Tare resetting Cancel or return to previous menu...

Страница 13: ...9 MENU MAP Into menus changes are applied right after pressing the key no further confirmation is required SETPOINT SYSTEM PARAMETERS 3 s...

Страница 14: ...10 3 s TLB4 only...

Страница 15: ...the indicated weight if necessary see section REAL CALIBRATION WITH SAMPLE WEIGHTS If the instrument HAS NOT BEEN CALIBRATED missing plant system identification tag proceed with calibration If the sy...

Страница 16: ...a value between 0 50000 and 7 00000 mV V Example of 4 cell system with sensitivity 2 00100 2 00150 2 00200 2 00250 enter 2 00175 calculated as 2 00100 2 00150 2 00200 2 00250 4 the division resolution...

Страница 17: ...ght is set to zero the value is stored to the permanent memory Press to display the value of the total weight reset by the instrument given by the sum of all of the previous zero settings DIAGNOSTICS...

Страница 18: ...this phase all of the LEDs are flashing Confirming once again the weight is set to zero the value is stored to the permanent memory Press to display the value of the total weight reset by the instrume...

Страница 19: ...ght too the system must read zero If this does not happen it means that there is a mechanical problem affecting the system linearity WARNING identify and correct any mechanical problems before repeati...

Страница 20: ...nt the display shows the automatically detected channels press to exit with no change press to make the current selection permanent and disable automatic detection In case a channel is not working you...

Страница 21: ...load cell 2 wait for stability and confirm by pressing Repeat the operation for each connected load cell If equalization is successfully completed the display shows confirm by pressing to exit if an e...

Страница 22: ...ter enables to stabilise a weight as long as its variations are smaller than the corresponding response time It is necessary to set this filter according to the type of application and to the full sca...

Страница 23: ...on the weight value is lower than the value set in this parameter and does not exceed the value the weight is reset To disable this function set 0 ZERO TRACKING from 1 to 5 default When the weight va...

Страница 24: ...nt parameter see the related section To use is necessary to enable it closing the input see section OUTPUTS AND INPUTS CONFIGURATION DISPLAY COEFFICIENT By setting the coefficient the display is chang...

Страница 25: ...are expressed in weight value If you want to convert them to the new unit of measurement perform one of the following procedures for changing the system calibration The parameter must remain set to 1...

Страница 26: ...operation mode is selected the following options are also active the contact will switch on the basis of gross weight the contact will switch on the basis of net weight If the net function is not acti...

Страница 27: ...etting of the units of measure and coefficient otherwise the weight is displayed when the input is closed the data are sent for printing if in the communication protocol of either serial port the para...

Страница 28: ...t is displayed keep pressed to display the gross weight When the key is released the net weight will be displayed again IF A SEMI AUTOMATIC TARE NET IS ENTERED IT IS NOT POSSIBLE TO ACCESS THE ENTER P...

Страница 29: ...a drop of tin choice of a weight followed by the analog output gross or net If the net function is not active the analog output varies according to gross weight set the weight value for which you wis...

Страница 30: ...NALOG OUTPUT TYPE Minimum Maximum 0 10 V 0 150 10 200 0 5 V 0 150 5 500 10 V 10 300 10 200 5 V 5 500 5 500 0 20 mA 0 200 22 000 4 20 mA 0 200 22 000 NOTE the analog output may also be used in the oppo...

Страница 31: ...t continuous weight transmission protocol to RIP675 RIP6125C series remote displays set When the remote display is set to gross weight if the instrument displays the gross weight the remote display sh...

Страница 32: ...CATION If the RS485 network exceeds 100 metres in length or baud rate over 9600 are used two terminating resistors are needed at the ends of the network close the two jumpers indicated in the picture...

Страница 33: ...nd or channels in error it displays stored load distributions weight load percentage on each channel it deletes stored load distributions confirmation is requested press to proceed or press to cancel...

Страница 34: ...tus weight load percentage on each channel enabled channels and or channels in error Input Test ensure that for each open input is displayed is displayed when the input is closed Output Test setting e...

Страница 35: ...le weight used tare setting via the keypad press to display the set value load distribution error press to display the weight value that triggered the alarm press again to display the difference betwe...

Страница 36: ...urs when the weight exceed the value set in this parameter The type of switching is settable see section OUTPUTS AND INPUTS CONFIGURATION from 0 to max full scale default 0 Hysteresis value to be subt...

Страница 37: ...cted to the multi channel weight transmitter which transmits the weight to the indicator all the operations of calibration zeroing equalization and channel selection can be performed remotely through...

Страница 38: ...ING the map shows only the additional menu items that are enabled on the W series instruments connected to the TLB4 Into menus changes are applied right after pressing the ENTER key no further confirm...

Страница 39: ...cates exactly what would be displayed on the TLB4 display If the configuration is performed on the TLB4 the W series instrument connected must be restarted to allow synchronization REMOTE KEYPAD LOCKI...

Страница 40: ...g in real calibration after the fifth sample weight value has been entered the value set for the parameter is beyond the permitted values press to quit the setting mode leaving the previous value unch...

Страница 41: ...II __O F_ __O L_ __O F_ __O L_ __O F_ aa CR RIP __O F_ __O L_ __O F_ __O L_ __O F_ __O F_ HDRIP N _ERCEL _ER_OL _ER_AD _ER_OF O__SET CONTIN _ERCEL _ER_OL _ER_AD _ER_OF O__SET For RIP remote displays i...

Страница 42: ...NGS from the weight display press the key BASIC PRINTOUT TLB4 BASE Addr 01 GROSS 878 kg NET 589 kg TARE 289 kg BASIC PRINTOUT PEAK ENABLED TLB4 BASE Addr 01 GROSS 1204 kg NET 831 kg TARE 373 kg PEAK 2...

Страница 43: ...ION TLB4 BASE Addr 01 CURRENT STATUS GROSS 2014 kg CH1 23 5 CH2 24 1 CH3 ERR load cell connected but in error CH4 OFF load cell not connected CURRENT AND STORED DISTRIBUTION TLB4 BASE Addr 01 STORED M...

Страница 44: ...ncluded mounting kit drilling size 23x96 mm maximum panel thickness 2 5 mm Pull out the terminal strips and insert the instrument into the hole Insert the mounting brackets into the housings on both s...

Страница 45: ...ff to indicate that this menu item is unlocked TEMPORARY MENU UNLOCKING press and simultaneously for 3 seconds it is now possible to enter and modify all menus including those which are locked By retu...

Страница 46: ...the proper manual and the correct procedures for approval indicating mandatory hardware code and serial number see section INSTRUMENT COMMISSIONING When the TLB4 is used in combination with a W serie...