- 7 -
TECHNICAL SPECIFICATIONS
POWER SUPPLY and CONSUMPTION
12/24 VDC ±10%; 5 W
No. OF LOAD CELLS IN PARALLEL and SUPPLY
max 4 (350 ohm); 3.3 VDC / 40 mA
LINEARITY / ANALOG OUTPUT LINEARITY
< 0.01% F.S.; < 0.01% F.S.
THERMAL DRIFT / ANALOG OUTPUT THERMAL DRIFT < 0.0005% F.S./°C; < 0.003% F.S./°C
A/D CONVERTER
A/D CONVERTER
MAX DIVISIONS
(with measurement range ±6 mV = sens. 2 mV/V)
±999999; 6.6 nV/d
MEASUREMENT RANGE
±26 mV
MAX SENSITIVITY OF USABLE LOAD CELLS
±7 mV/V
MAX CONVERSIONS PER SECOND
600 conversions/second
DISPLAY RANGE
±999999
No. OF DECIMALS / DISPLAY INCREMENTS
0÷4 / x 1 x 2 x 5 x 10 x 20 x 50 x 100
DIGITAL FILTER / READINGS PER SECOND
10 levels / 5÷600 Hz
RELAY OUTPUTS
3 - max 115 VAC; 150 mA
24 VDC; 200 mA
DIGITAL INPUTS
2 - 5÷24 VDC PNP
MICRO USB PORT
type B – USB 2.0 (full speed)
HUMIDITY (non condensing)
85%
PROTECTION RATING
IP67
STORAGE TEMPERATURE
-30°C +80°C
WORKING TEMPERATURE
-20°C +50°C
DIMENSIONS
97x38x82 mm (including flying connectors)