Work prior to commissioning
Part 2 • Page 3-3
Part 2 – Supplementary Operating Instructions
Rev. 4.00 December 2014
Software version 1.04
EcoRO Dia I/II (HT)
Part 2 • Chapt
er 3
3.1.3
Electrical connection
•
The EcoRO Dia I/II (HT) requires pre-fuses, ground fault protector (FI
∆
N
30mA) and an external power
switch according the following performance data’s
•
Observe backup fuse requirements on-site in accordance with regulations.
•
4x shock-/waterproof socket outlet (115V) protected by ground fault protector FI
∆
N
30mA.
•
Alarm cable 7x AWG18 (1.0mm²) (option)
•
Remote control cable 7x AWG18 (1.0mm²) (option)
•
Ethernet port RJ45 for Aquaboss Vision
3.1.4
Permanent system connection
•
Overcurrent protection devices must be installed in the building installation
Circuit breaker
Ground fault protector => 30 mA / 4-pole
•
Device or power switch 63 A
Device type
Voltage
Capacity
Current
EcoRO Dia II (HT) 700
3~N PE 208V
11.54 kVA
31 A
EcoRO Dia II (HT) 900
3~N PE 208V
11.54 kVA
31 A
EcoRO Dia II (HT) 1050
3~N PE 208V
14.92 kVA
41 A
EcoRO Dia II (HT) 1400
3~N PE 208V
13.19 kVA
36 A
EcoRO Dia II (HT) 1800
3~N PE 208V
14.92 kVA
41 A
EcoRO Dia II (HT) 2100
3~N PE 208V
16.58 kVA
45 A
EcoRO Dia II (HT) 2450
3~N PE 208V
16.58 kVA
45 A
EcoRO Dia II (HT) 2800
3~N PE 208V
16.58 kVA
45 A
EcoRO Dia II (HT) 3600
3~N PE 208V
16.58 kVA
45 A
WARNING
It is not recommended that the R.O. unit operate on an emergency eletrical supply or any
unstable power supply that may over-voltage or under-voltage the water treatment system.
Protection against electrical surges is recommanded.
NOTICE
These operating instructions covers the 208 VAC/60 HZ model of the
Aquaboss
®
EcoRO
Dia I/II (HT) water treatment system. Always confirm voltages and power supplies before
operating the unit.
NOTICE
The separator must meet the requirements in accordance with IEC 60947-1 and IEC 60947-3.
•
The switch must not interrupt the earthing conductor.
•
The function of the switch as separator of the RO from the mains must be be marked
by a label.
•
The switch must be near the device and be easily accessible for the operator.
NOTICE
When a permanent mains cable is attached, it has to meet the requirements according to
DIN EN 61010-1/6.10.2.