7.9

 

Ramp function 

With the ramp function, temperature changes over any time period can be conveniently entered. This is especially 
advantageous with very low temperature changes (e.g. 0.1 °C/ day).  
Example: From the current outflow temperature (e.g. 242.4 °C) 200 °C of cooling is to occur over 5 days.  
Then 200 °C is entered as the temperature change, the time value 5 is entered for the time and day(s) selected as the 
time unit. 

 

The ramp function is executed until it is manually terminated or until the temperature limits T

il

 (min) or 

T

ih

 (max) described in Section 7.6.2 are attained. 

 

Command 

 

 

 Ramp function 

 

 

 

Open the list of device parameters using the 

soft key    Menu . 

 

With the cursor keys continue to  

 

 Programmer 

 

 

 Ramp function 

 

Confirm with the key 

 

 

Enter a positive or negative temperature value 
with  Temp. change . 

 

With  Time  enter a figure (without time unit).  

 

With  Time unit  choose 
between 

 Second(s) 

 up to 

 Day(s) 

 

Under  Status  the ramp is started  

 

 Start 

 or stopped 

 

 Stop 

 

When the ramp function is being 
executed, 

 Ramp active 

 appears in the window 

bar. 

 

Without manual switch-off the ramp 
terminates at the latest at T

il

 (min) or T

ih

 (max). 

 

 

 Pump  Menu 

End 

T

set

 

T

fix

 

Pump 
Settings 
Graph 
Clock 

Programmer          

Interfaces 
Control 
Limits 

 

Program 1        

 

Program 2 
Program 3  
Program 4  
Program 5  

Ramp function        

 

 

 

 Pump  Menu 

End 

T

set

 

T

fix

 

Status 
Temp. change 
Time 

Time unit               

 

Second(s)           

 

Minute(s) 
Hour(s) 

Day(s)                    

 

 

Содержание RP 4050 C

Страница 1: ...ation manual PROLINE Kryomats Low temperature thermostats with SmartCool System RP 4050 C RP 4050 CW RP 3090 C RP 3090 CW RP 4090 C RP 4090 CW Release 03 2022 j Read the instructions before starting w...

Страница 2: ...ces release 08 2020 i 11 2018 e 10 2018 d 05 2016 b2 03 13 01 09 06 08 Valid from software control system Master version 2 13 software safety system Master version 2 07 software operating system Comma...

Страница 3: ...air moving the equipment Transport the equipment with care The unit may NEVER be overturned nor put upside down Equipment and its internal parts can be damaged by dropping by shock Technically qualifi...

Страница 4: ...MODULES ACCESSORIES 20 4 9 REFRIGERATING UNIT 21 4 10 AVOIDANCE OF DEWING 21 4 11 HEATER RATING AND POWER CONSUMPTION FROM THE MAINS 21 5 UNPACKING 22 5 1 AFTER UNPACKING 22 5 2 STANDARD ACCESSORIES 2...

Страница 5: ...he offset of the external temperature probe 64 7 6 9 Restoring the works setting of the external temperature probe offset 64 7 7 GRAPHICAL DISPLAY OF TEMPERATURE MEASUREMENTS 65 7 8 PROGRAMMER 67 7 8...

Страница 6: ...NTACT MODULE 108 8 5 1 Contact module LRZ 915 with three inputs and three outputs 108 8 5 2 Namur Contact module LRZ 914 with only one input and one output 109 9 MAINTENANCE 110 9 1 DEVICE STATUS 110...

Страница 7: ...2 Proline Kryomats 7 137 9 5 HELP DESK AND ORDERING REPLACEMENT PARTS 119 10 ACCESSORIES 120 11 TECHNICAL DATA AND DIAGRAMS 121 12 DECLARATION OF CONFORMITY AND PRODUCT RETURNS DECLARATION 130 13 INDE...

Страница 8: ...quences of non compliance Action 1 Action WARNING indicates a possible dangerous situation which if the safety requirements are ignored may result in fatal or severe irreversible injuries Caution Type...

Страница 9: ...le 2 Industrial in accordance with DIN EN 61326 1 Emissions Class B in accordance with CISPR 11 Only for EU Domestic connection value 100 A Proline Kryomat Table 2 Industrial in accordance with DIN EN...

Страница 10: ...thermal transfer liquid for temperatures above 100 C carefully remove all water residues including from the hoses and loads When doing this also open the blanking caps of the pump outputs and inputs...

Страница 11: ...IN EN 61010 2 010 Take care of the level of the heat transfer liquid 6 2 3 Compare the information on the rating label with the supply details 4 Connect the unit only to a socket with a protective ear...

Страница 12: ...T End End C Control parameters setpoint resolution setpoint source input calibration etc Protection system Command cooling system analog module RS 232 module and digital module etc Display flashes Set...

Страница 13: ...trast intern Pt100 extern Pt100 Device type Software Version Serial numbers Device data Error store Mode Baud rate RS 485 address internal external Pt100 Analog interface RS 232 485 Calibration Defaul...

Страница 14: ...illustration on this page 4 Bath cover 5 Rotary switch for power supply 6 Grille on both sides 7 Front cover closed 8 Four steering transport rollers two off them with stoppers 1 Pump connection on th...

Страница 15: ...n tap 2 Bath drain nozzle 1 Bath edge heating and bath bridge heating 2 Nameplate 3 Mains cable 4 Connection cable for control head 5 Connection cable LiBus for control head 6 Connection bath bridge h...

Страница 16: ...t LiBus LAUDA internal bus for bus suitable for unit and to which the refrigerating lower section and Command remote control are connected 5 Mains switch 6 Connection socket 51H 7 Air intake for elect...

Страница 17: ...gnal red LED is blinking 9 Overtemperature set point to check or set Tmax 10 Mains On green LED is lit Control element Command 1 Graphical display 2 RS 232 485 socket hidden on the back of Command 3 C...

Страница 18: ...ble drive pump level 5 to 8 The pump power can therefore be optimally matched to the relevant task In order to achieve optimum temperature homogeneity in the heat transfer liquid in the bath it is rec...

Страница 19: ...e in the bath A high resolution A D converter processes the measurement Further measurement conditioning occurs using a special control algorithm for controlling the heater actuator which has a low re...

Страница 20: ...nterfaces The device is equipped in series with the following sockets One socket 10S for the connection of an external Pt100 temperature probe Two sockets 70S for the connection of components via the...

Страница 21: ...ts must not be restricted respectively covered 6 1 The Proline Kryomats are equipped with the SmartCool technology which makes optimum use of the compressor and only then cools when refrigerating capa...

Страница 22: ...a level surface Observe the packing instructions Notice Falling toppling equipment Property damage Do not tilt the cooling device during transport and never turn it upside down To repack the unit car...

Страница 23: ...crew plugs for M16 x 1 already mounted for all Kryomats EOA 001 2 Threaded house coupling Nipple Nut G for water cooled devices only EZB 260 1 Warning label Hot surface for all Kryomats 5 3 Packing an...

Страница 24: ...022 5 3 3 Unpacking the device 5 3 3 1 Outer cardboard box Open the cardboard box on top Pallet Take out the filling material and the accessory parts Draw out the four nails on the edge which fix the...

Страница 25: ...e the outer cardboard box vertically upwards 5 3 3 2 Lift device from the pallet Two textile round slings Attention The round slings must not press against the structure of the device Pay attention to...

Страница 26: ...device length Textile round slings 5 3 4 Repacking with original packing material The packaging of a device takes place in the reverse order 5 4 In plant transport with hand pellet truck or fork lifte...

Страница 27: ...ts 27 137 Transport the device on a hand pellet truck Transport the device with a fork lifter 5 5 Before preparation Remove the protective foil For installation and commissioning the operating instruc...

Страница 28: ...the device Crushing of the hands and feet Do not place any heavy parts on the device Caution Contact with hot cold heat transfer liquid Scalds frost bite Bring the heat transfer liquid to room temper...

Страница 29: ...Master control element and secure it Operation with external loads Circulating thermostat continue at 0 Check whether the pump connectors at the side and back are fitted with sealing caps 3 or that h...

Страница 30: ...splay of the warning WXArn103 7 12 4 Changes in volume of the heat transfer liquid during temperature changes should be taken into account Best operation is with a level 30 80 mm below the top edge of...

Страница 31: ...e mains switch rotary switch withdraw the mains plug The drain tap 1 is located behind the front panel Fit the hose onto the draining nozzle 2 Open the drain tap and run off the heat transfer liquid C...

Страница 32: ...B 214 LZB 314 Therm 250 50 250 Silicone oil 130 45 at 50 C 300 LZB 122 LZB 222 LZB 322 At higher temperatures Evaporation losses Use bath covers Only use distilled water or fully demineralized high pu...

Страница 33: ...rantee Important Danger of corrosion of the cooling water circuit due to water of unsuitable quality Free chlorine e g from disinfectants and water containing chlorine lead to pitting in the cooling w...

Страница 34: ...ollution due to leakage To avoid pollution due to a leak in the cooling water system it is recommended that a leakage water detector with a water cut off is installed Servicing intervals Follow the in...

Страница 35: ...hazard from leaking heat transfer liquid Fit sealing plugs to the pump connections when no external consumers are connected and set the flow distribution to internal INT Operation as circulating therm...

Страница 36: ...ning the union nuts on the pump nipple AF 19 use a wrench AF 14 to counter the tightening torque see figure If external control is to be used provide a Pt100 temperature probe in the external load 7 5...

Страница 37: ...ith a hydrometer Warning Risk of heat transfer liquid overheating due to incorrect entry of overtemperature switch off point Tmax Fire The overtemperature switch off point must be below the flash poin...

Страница 38: ...ntact with vapors from the heat transfer liquid Harmful by inhalation Use an extractor hood If possible use a bath cover Caution Bath overflow due to thermal expansion or immersion of objects Burns sc...

Страница 39: ...a 16 ampere circuit breaker fitted during installation Exception Devices with 13 ampere UK plugs Three phase devices For three phase devices the rating of the circuit breaker must match the power cons...

Страница 40: ...at transfer liquid this has been pumped out by filling the external load Display for LEUEL low level appears when the bath has too little liquid Red LED above the fault triangle flashes Find cause of...

Страница 41: ...g display remains active so that status displays can be viewed and settings carried out However a started timer 7 10 continues to run Stop as required with Pause 7 4 Key functions Your Proline Thermos...

Страница 42: ...onnected The following always applies After termination of the relevant settings they are accepted automatically after approx 4 s or the setting is accepted immediately with the Enter key and Fault si...

Страница 43: ...e to change the setpoint temperature then press the duo key under Tset Brightness Contrast The brightness and contrast can be set on the Command remote control The works setting can be changed via Set...

Страница 44: ...normal super and graphics recorder windows Tset Changes setpoint temperature Tfix Calling and setting of saved setpoints 2 Standard window with five important items of information Tint current bath te...

Страница 45: ...shown graphically against time 7 7 Soft keys like above 7 4 2 Changing window information Command Display data You can adapt the information displayed by your Command remote control to your requiremen...

Страница 46: ...vantageous when the thermostat is positioned in another room and the Command remote control is used as a remote control device Then the Master keyboard can be locked to prevent unintentional adjustmen...

Страница 47: ...pressed simultaneously for three seconds The locking window appears Hold both keys pressed until the progress bar is completely filled Then the display skips back to the previously set Screen mode The...

Страница 48: ...ait 4 seconds or Display flashes 4 seconds new value is automatically accepted or value is accepted immediately with Enter key For safety reasons the setpoint can only be set up to 2 C above upper lim...

Страница 49: ...h or Confirm the value with or quit the window with without having made any changes Two other ways of entering the setpoint With the soft key Tfix open the window shown on the left The setpoints which...

Страница 50: ...can be used as the controlled variable for the bath temperature in applications with a noticeable temperature gradient between the internal bath temperature and an external load The setup is described...

Страница 51: ...modules 8 7 5 3 Setting pump power or standby With the Proline Varioflex pump four pump levels level five till level eight are available with which the bath circulation flow rate and pressure the nois...

Страница 52: ...ow with without making any changes It is not possible to select the pump levels 1 to 4 Standby activation Standby activation Pump heater and refrigerating machine are deactivated when the green LED in...

Страница 53: ...ny changes Automatic mode means that the external pump is switched off while the unit is in standby mode or it is switched on while the unit is running 7 5 5 Activating external control An external te...

Страница 54: ...ource 1for the internal probe EPonly when an external probe is connected EAonly when an analog module is connected and configured ESonly when a serial module is connected and is continuously receiving...

Страница 55: ...the soft key Menu With the cursor keys change further to Settings Basic settings Current Consumpt 16 0 A is active Open the settings window with Changes are not accepted Quit the window with Pump Men...

Страница 56: ...Change the time with cursor or soft keys and accept with or quit the window with without making changes The date is set just the same with Set date The date format Day Month Year or Month Day Year ca...

Страница 57: ...key Menu With the cursor keys continue to Settings Display resolution Select the desired resolution with or Accept selection with or End or quit the window with without making any changes Pump Menu E...

Страница 58: ...art Open the device parameter menu via the soft key Menu With the cursor keys continue to Settings Basic settings Auto start On is currently active If the standby mode is to be activated after a mains...

Страница 59: ...the maximum temperature and 5 C as the minimum temperature Command Limits Open the device parameter menu via the soft key Menu With the cursor keys continue to Limits The minimum and maximum temperat...

Страница 60: ...s continue to Control Setpoint offset Offset source Off indicates that the setpoint offset is currently deactivated Select the setpoint source with or and confirm with Interfaces e g RS 232 are only d...

Страница 61: ...orks setting with all default Under Master you have the choice between all default then all Master settings are reset only control para int for the internal control parameters only control para ext si...

Страница 62: ...ering the offset of the internal temperature probe If during checking with a calibrated reference thermometer a deviation is found then the offset i e the additive part of the characteristic of the in...

Страница 63: ...onally the works setting can be restored with this function Command Default Open the device parameter menu via the soft key Menu With the cursor keys continue to Settings Calibration intern Pt100 Defa...

Страница 64: ...eys continue to Settings Calibration extern Pt100 Calibration The adjacent window appears Confirm selection with Continue as described in 7 6 6 for the internal temperature probe 7 6 9 Restoring the w...

Страница 65: ...is to be invisible or visible Sample time defines with which time interval the measurements are recorded 5 possibilities are offered From 2s max 1h45min up to 2min max 105h Time axis defines over whi...

Страница 66: ...ter the desired new value in the changes window in the usual way When setting the minimum value the largest permissible value here 26 90 C since the maximum value is 27 C is stated When setting the ma...

Страница 67: ...ibed by the target temperature i e the temperature at the end of the segment and the duration from the start to the end of the segment Step Without any specified time the final temperature is approach...

Страница 68: ...ction 0 The field tolerance refer to the above program table and the graph below It facilitates exact conformance to the dwell time at a specified temperature Segment 1 is not processed until the bath...

Страница 69: ...bmenu Status appears Using the Status menu the selected program can be 1 started Start 2 paused Hold 3 continued Continue or 4 terminated Stop In addition the standby key stops the programmer Pause op...

Страница 70: ...r Stop are shown Here with the aid of the keys or and the running program can be paused with Hold or terminated with Stop Once the program has been terminated the device runs with the last setpoint se...

Страница 71: ...letion of a segment In addition when a program has just been executed new segments can be inserted and existing ones modified even the currently active segment Furthermore all segments except the curr...

Страница 72: ...hange It can be edited by pressing the key see following pages The soft key Insert inserts in the marked line a new segment that has a default value taken from the previous segment with the exception...

Страница 73: ...pied If the field in the column T end C has a black background the entry mode End of segment temperature is obtained by pressing the key Depending on the setting that is the temperature which the ther...

Страница 74: ...m with Continue with to the entry field Pump If the field in the column Pump has a black background the entry mode for the Pump level is obtained by pressing the key With or select Pump Level 5 8 or a...

Страница 75: ...Contact output is obtained by pressing the key With or select Open or Closed and confirm with stands for no change with respect to the previous segment i e if is present in all fields the contact set...

Страница 76: ...gram to repeat continuously Confirm the entry with the key and return to the display You can quit the Edit mode with or End 7 8 6 Viewing the program sequence as a graph Graph Command Programmer Progr...

Страница 77: ...necessary to process step changes in temperature Number of free segments Segment which is at present currently being processed Residual time of the current segment in hours and minutes Current pass i...

Страница 78: ...re attained Command Ramp function Open the list of device parameters using the soft key Menu With the cursor keys continue to Programmer Ramp function Confirm with the key Enter a positive or negative...

Страница 79: ...n from the standby mode The menu Function is used to define when an action is executed Similar to an electronic mains timer Weekplan enables two switching events to be carried out each day The cycle i...

Страница 80: ...bath thermostat with water as the bath medium with internal control The parameters are also preset for the operation of external containers with external control Sometimes however the operation of ext...

Страница 81: ...ater 9 0 60 42 6 3 8 RP 4090 CW Ethanol 9 0 60 42 6 3 8 RP 4090 C Water 9 0 60 42 6 3 8 RP 4090 C Ethanol 9 0 60 42 6 3 8 Technical changes reserved 7 11 2 External control variable External measureme...

Страница 82: ...t the parameters to be changed with and confirm with Then change the value in the following settings window and confirm with Correction limitation see introduction 7 11 2 1 Steps for setting the contr...

Страница 83: ...oscillations with large period i e 10 min Kpe higher otherwise lower 3 2 Tne Tve Tde setting Start with high numbers Tne 70s 200s Tve 50s 150s With lower numbers faster approach otherwise slower appro...

Страница 84: ...later reuse With Default the set of control parameters saved at the works is loaded again in this case the control parameters set by the customer are lost Command Control parameter sets Open the devic...

Страница 85: ...run of the thermostat In this case the thermostat and if applicable the external application must be ready for operation 6 The Self Adaption will be performed with the actually set pump step Best res...

Страница 86: ...cillate less than about 15 Kelvin around the set temperature Change the display in the adjacent window and accept with With the menu Identification the optimal control parameters for internal control...

Страница 87: ...parameter set 9 After the test run the detected control parameters will be taken over as control parameters automatically If the parameters found do not fulfil your expectations the before set paramet...

Страница 88: ...ce please give information to the LAUDA Service 9 5 Find cause of alarm or warning and rectify where necessary Then press on the Master keyboard in order to remove the alarm message Warning messages c...

Страница 89: ...intended to prevent unintentional adjustment Set the overtemperature switch off point Tmax below the flash point of the heat transfer liquid The setting range is restricted to 5 C above the upper lim...

Страница 90: ...ter switches off on both poles Varioflex pump and refrigerating unit are switched off Find the cause of the fault and where necessary top up for missing liquid 6 2 and 6 3 Press the Enter key Also pre...

Страница 91: ...el goes down This is the factory setting Warning and heater off nHWXH Warning heater off Warning and additional heater off as long as the level goes down Recommended for flammable heat transfer liquid...

Страница 92: ...s Heater switches off on both poles Pump and refrigerating unit are switched off electronically Find the cause of the fault Possible causes may be 1 Volume expansion on heating 2 Feed to an external v...

Страница 93: ...Master control panel 7 12 6 Pump motor supervision Dry running The SelfCheck Assistant monitors the Varioflex pump 1 Alarm sounds as dual tone signal when the pump runs without liquid This can only o...

Страница 94: ...ompressor has time enough to cool down again This error may affect stage 1 Error 68 or stage 2 Error 69 If the error is always present after switching on the device please give information to the LAUD...

Страница 95: ...4 tih limitation active WX104 Bath level is approaching switch off level or is out of optional range WX 5 Heatsink temperature is superheated WX105 Heater 1 break WX106 Heater 2 break WX107 Heater 3 b...

Страница 96: ...Software version of analogue interface too old WX215 Software version of RS 232 too old WX315 Software version of RS 232 too old WX216 Software version of contact I 0 too old WX316 Software version of...

Страница 97: ...XX Message Meaning Message Meaning WX601 Overflow of CAN receipt WX701 Overflow of CAN receipt WX602 Watchdog Reset WX702 Watchdog Reset WX610 Software version of control system too old WX710 Software...

Страница 98: ...the back of the master control head into two module slots Switch off the Kryomat at the rotary switch on the front panel and withdraw the mains plug Touch the earthed bath cover of the Proline thermo...

Страница 99: ...ting signal Pump power Pump speed 2400 4800 9600 19200 Mode Baud rate RS 485 address Voltage 0 10V Current 0 20mA Current 4 20mA Status Function Interface type Minimum value Maximum value Calibration...

Страница 100: ...ostat to the PC use 1 1 cable and not a null modem cable without hardware handshake the computer PC must be set to the operating mode without hard ware handshake Use screened connecting cable Connect...

Страница 101: ...command This ensures that the sequencing of inquiries and answers is clear CR Carriage Return Hex 0D LF Line Feed Hex 0A Example Transfer of setpoint 30 5 C to the thermostat Computer Thermostat OUT_S...

Страница 102: ...X External temperature to be set through the interface OUT_SP_00_XXX XX Setpoint transfer with up to 3 places before the decimal point and up to 2 places behind OUT_SP_01_XXX Pump output step 1 to 8 O...

Страница 103: ...ter cooling unit OFF RMP_SELECT_X Selection of the program 1 5 to which the further instructions apply When the unit is switched on program 5 is selected automatically RMP_START Start the programmer R...

Страница 104: ...R_05 Read control parameter TnE 980 OFF IN_PAR_06 Read control parameter TvE 0 OFF IN_PAR_07 Read control parameter TdE IN_PAR_09 Read value of correction limitation IN_PAR_10 Read control parameter X...

Страница 105: ...t adjustment alarm 7 char no external control variable RMP_IN_00_XXX Read a program segment XXX response e g 030 00_010 00_005 00_001 00 set point temperature 30 00 C time 10 min tolerance 5 00 C pump...

Страница 106: ...Syntax error in value ERR_6 Illegal value ERR_8 Module ext temperature not available ERR_30 Programmer all segments occupied ERR_31 Set point not possible analogue set point input ON ERR_32 TiH TiL E...

Страница 107: ...d Set temperature The temperature source with which active control occurs MN tC Controlled temp Actual temperature bath temperature MN t1 or Internal Temp External actual temperature from Pt100 MNtEP...

Страница 108: ...tb or Stand by Control programmer Input 1 activates programmer 1 input 2 activates programmer 2 etc At the first close the programmer gets starting open removes it in pause The next close initiate con...

Страница 109: ...and one input on each of two DIN sockets Contact inputs and outputs Output Input View on flange plug Front or solder side coupler socket Max 30 V 0 2 A Coupler socket Catalogue number EQD 047 View on...

Страница 110: ...splayed 9 1 4 Device data Command Device data Settings Device status Device data Display Text shows various actual temperatures in C from ext Pt100 and the modules T cont head and T heatsink are tempe...

Страница 111: ...ard Disconnect the device from the mains before cleaning Notice Live parts in contact with cleaning agents Property damage Disconnect the device from the mains before cleaning Water and other liquids...

Страница 112: ...the water The display changes automatically to the overview window To close the valve again selects this menu again The item is called now End unfill conden 9 3 Servicing cleaning repair and disposal...

Страница 113: ...ed out only by a specialist 9 3 1 Servicing LAUDA Thermostats largely require no service If the heat transfer liquid becomes contaminated it should be replaced 6 2 At the back of the Master head a mai...

Страница 114: ...ture protection Quarterly 7 12 1 Low level alarm warning Quarterly 7 12 2 9 3 3 Testing the heat transfer liquid Caution Contact with hot cold heat transfer liquid Scalds frostbite Bring the heat tran...

Страница 115: ...dust at intervals of one month or longer depending on the operating period and dust level in the ambient air To do this open the front grille brush off the condenser and where necessary blow over wit...

Страница 116: ...r inlet and between cooling water outlet and back to container Change containers when top one is empty Acting time Continue the pump stage until most of the foamy reaction usually at the start has dec...

Страница 117: ...ertemperature cut off point or set the cut off point higher than the outflow temperature Master Warning message WXArn 104 Command Level very low Imminent low level in the bath vessel Master Alarm mess...

Страница 118: ...or 11 Command Overpressure outflow pressure too high Pump level too high Select a lower pump level 7 5 3 Compressor overtemperature 7 12 7 Three phase current 7 12 8 9 4 Disposal information The follo...

Страница 119: ...rument type and serial number from the nameplate This avoids queries and supply of incorrect items Nameplate Your contact for service and support LAUDA Service Telephone 49 0 9343 503 350 English and...

Страница 120: ...pter cable for the LAUDA internal bus LiBus For the connection of further LiBus components with heating thermostats two LiBus connections are not occupied and one with cooling thermostats EKS 073 Exte...

Страница 121: ...Master 0 01 Command 0 1 0 01 0 001 Display accuracy 0 2 C can be calibrated additively Section 1 3 last Point Temperature stability 10 C with ethanol K 0 05 Safety equipment Class III FL suitable for...

Страница 122: ...L 32 44 Bath opening B x L mm 350 x 350 Bath depth usable depth mm 250 230 Height to top of bath mm 905 Overall dims B x L mm 600 x 700 Overall dim H mm 1160 Weight kg 130 Power consumption 400 V kW 5...

Страница 123: ...be calibrated additively Section 1 3 last Point Temperature stability 10 C with ethanol K 0 05 Safety equipment Class III FL suitable for flammable and non flammable liquids Cooling Air Water Air Wate...

Страница 124: ...x 1 olives 13 mm external diameter Bath volume from to L 23 31 32 44 Bath opening B x L mm 350 x 200 350 x 350 Bath depth usable depth mm 250 230 Height to top of bath mm 905 Overall dims B x L mm 600...

Страница 125: ...atalogue number L001669 L001670 L001671 L001673 L001674 L001675 L001693 L001694 L001695 L001698 L001699 L002583 Operating temperature ACC range C 40 150 ACC range is the working temperature range duri...

Страница 126: ...ogue number L001661 L001662 L001663 L001665 L001666 L001667 L001685 L001686 L001687 L001690 L001691 L003524 L003525 Operating temperature ACC range C 90 150 ACC range is the working temperature range...

Страница 127: ...The device contains fluorinated greenhouse gases Unit RP 4050 C RP 4050 CW Refrigerant R 449A R 449A maximum filling quantity kg 1 6 1 6 GWP 100a 1397 1397 CO2 equivalent t 2 2 2 2 Devices with two c...

Страница 128: ...128 137 Proline Kryomats 03 2022 Pump characteristics measured with water internal pump PL 4 Cooling curves L min Flow rate Pressure bar...

Страница 129: ...03 2022 Proline Kryomats 129 137 Cooling curves Bath closed Heat transfer liquid Ethanol Time in minutes Temperature in C Influence of ambient temperature at air cooled Kryomats...

Страница 130: ...130 137 Proline Kryomats 03 2022 12 Declaration of conformity and product returns declaration...

Страница 131: ...03 2022 Proline Kryomats 131 137...

Страница 132: ...43 E EA analog module control source 54 Edit 72 EMC standard DIN EN 61326 1 9 Enter key Command 42 Master 41 EP external probe control source 54 Error list 88 error store 111 Errors 88 ES serial modu...

Страница 133: ...ck Assistant 19 88 Self test 40 Serial interface RS 232 485 100 Serial numbers 110 Service contact 119 Service intervals 114 Servicing 112 Set date 56 SEt setpoint setting 48 Set time 56 Setpoint offs...

Страница 134: ...134 137 Proline Kryomats 03 2022...

Страница 135: ......

Страница 136: ...Manufacturer LAUDA DR R WOBSER GMBH CO KG Laudaplatz 1 97922 Lauda K nigshofen Telephone 49 0 9343 503 0 Fax 49 0 9343 503 222 E mail info lauda de Internet https www lauda de...

Отзывы: