92 / 137
Proline Kryomats
7.12.4
High-level warning or alarm
3 Sec.
Acoustic warning signal sounds for 3 seconds when the liquid level rises so far
that the uppermost switching point of the level sensor has been reached.
Or in case the warning function as described in 7.12.3 was chosen:
The acoustic signal with dual-tone sounds.
Warning
WXArn103
(high level) appears when the bath contains too
much liquid.
The
WXArn
flashes by turns with the numeral.
In case the alarm function as described in 7.12.3 was chosen:
The acoustic signal with dual-tone sounds.
The red LED
above the fault triangle
flashes.
Heater switches off on both poles,
Pump and refrigerating unit are switched off electronically.
Find the cause of the fault. Possible causes may be:
1.
Volume expansion on heating.
2.
Feed to an external vessel may be interrupted so that only return suction
is possible.
3.
Heat transfer liquid taking up moisture.
If Alarm: Press Enter key. Warnings disappear automatically when the cause is
gone.
Also, press this key if the unit has been switched off in the fault state.
Warnings disappear automatically when the cause is gone.
Command
High-level warning/alarm
The display shows
Warning. To release press Enter key
Security 3 Level too high
or
Alarm AL 6: Level too high is shown in the display and signifies that
unlocking is only possible on the Master control panel.
Level warning
WX103
°C
Level alarm
AL 6
°C