Proline Kryomats
YAUE0007 / 10/2/2018
Starting up
87
7.11.4
Self Adaption
The function Self Adaption can be used to detect automatically the optimal control parameters for
internal or external control.
The Self Adaption can only be performed on a device with active cooling.
This function is available from software version 2.18 of Command. For thermostats with an older
software version a software update is necessary.
The Self Adaption determines the parameters by a test run of the thermostat. In this case the
thermostat and, if applicable, the external application must be ready for operation. (
6).
The Self Adaption will be performed with the actually set pump step. Best results can be achieved
with high pump steps.
The test run must be performed at a passive system; this means that during the test run an
exothermic
or endotherrmic reaction mustn’t take place.
The test run takes depending on the external application about 30 minutes to 3 hours. The bath
temperature will oscillate in this time less than about ±15 Kelvin around the set temperature. After the
test run the detected control parameters will be taken over as control parameters automatically.
Command
Self Adaption
Open the device parameter menu via
the soft key Menu .
With the cursor keys continue to:
Control
Control Parameters
Self Adaption
Settings
.
Confirm selection with
.
The window shown adjacent appears.
With the menu
Status
the test run of
the Self Adaption can be started.
When the Self Adaption is finished, the
test run will be terminated
automatically.
As soon as start
is pressed, in the
sofkey area the information
Adaption on
will be displayed
followed by the actual status of the test
run.
Pump Menu
End
T
set
T
fix
Control Parameters
Control para. sets
Tv manual/auto
Self Adaption
Correction limitation
Settings
Pump Menu
End
T
set
T
fix
Status
Setpoint
Identification
Actual Parameters
Start
Содержание RP 3050 C
Страница 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Страница 2: ......
Страница 4: ......
Страница 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Страница 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Страница 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Страница 134: ......
Страница 135: ......