Proline Low-temperature Thermostats
YACE0072 / 21.08.07
Starting up
27
7.3
Switching off / standby
Switching off:
Set mains switch to position 0. With operation at temperatures below 0° Celsius only
switch off the device via standby, because otherwise severe condensation and dew formation can
occur.
Standby operation:
Use the key
on the command console or by selecting the pump level Zero on
the master display. The pump, heating and cooling unit are switched off, but the operating display and
the electrical bath bridge heating for RP 890 and RP 1290 remain active, so that dew formation and
condensation into the bath continue to be suppressed.
However, a started timer
Ì
7.12 continues to run. Stop as required with Pause.
7.4 Key
functions
Your Proline Thermostat is easy to operate. For the Command Version you will most probably rarely
need to use these operating instructions.
7.4.1 General key functions and pilot lamps
Master
Enter key:
−
From the actual-value display at the main menu level,
−
activates input, display flashes,
−
saves input, display ceases to flash and menu point is left,
−
press for approx. 3 seconds: Exit function and returns to bath
temperature display.
or
−
Paging with keys is possible within the relevant level,
or setting of numerical values
Speeds up entry by moving the counting position to the left:
a) Keys are pressed and held down
or
b) one of the two keys is pressed and held down, followed
immediately by brief pressing of the other key.
Moves counting position to the right:
−
Switching
one
place to the right
occurs by briefly (1 second)
releasing the key, followed by another pressing of the key.
Useful additional information:
−
2 dots in the Master display indicate that a submenu follows.
−
3 dots in the display indicate that a submenu for a module
(interface...) or a component (thermostat, Command Console ....)
follows. Module/component-specific possible settings are only
displayed when the hardware is connected.
---