Proline Low-temperature Thermostats
YACE0072 / 21.08.07
Maintenance
115
9.2.4 Device
data
Master
Æ
LMDkt
Æ
RgmVW
Ì
−
Various device datas are displayed.
Command
Device data
Æ
Settings
Æ
Device status
Æ
Device data
Æ
Display
−
T ext shows various actual temperatures
in °C from ext. Pt100 and the modules.
−
T cont. head and T heatsink are
temperatures of electronics in the Master
in °C.
−
Pump power in Watts, speed in rpm,
current in ampere (A).
−
T
int
indicates the current internal bath
temperature in °C.
−
Mains voltage in percentage (%) of
nominal and frequency in hertz (Hz).
−
Level indicates the liquid level in the
internal bath.
−
Voltage of power transformer, 5V supply
and fan in Volt.
−
Cur. cons.: Mains current consumption in
Ampere.
9.2.5 Fault memory (Command)
For the analysis and localization of faults the Command version includes a fault memory in which up to
45 fault and alarm messages are saved.
Command
Error store
Æ
Settings
Æ
Device status
Æ
Error
store
Æ
Display
.
−
The last message is at the top.
−
Each message line can be marked with
the cursor keys. The message appears in
plain text in the footer.
−
Under Source the CAN node is displayed
which signaled the fault.
−
Code is the number, which in the Master
is shown in the display until the cause
has been rectified.
−
Type: Alarm, Warning or Fault (Error).
T ext Pt
25.70 Tint
-8
T ext analog
---.-- Mains U(%) 100.74
T ext serial
---.-- Mains frequ.
50
T cont. head 39.80 Level
4
T heatsink
51.68 Low voltage 27.90
Pump pow.
44.90 5V supply
5.00
Pump rpm
5460 Fan voltage
7.0
Pumpe cur.
1.68 Cur. cons.
2.84
Help Menu End T
set
T
fix
No. Source Code Type Date
Time
10 Safety 2 Alarm ------------ -----------
9 Safety 4 Warn. 28.08.03 15:32:02
8 Contro. 32 Error 17.07.0310.:52:02
7 Contro. 3 Warn. 06.06.03 11:15:11
6 Contro. 9 Alarm 05.06.03 08:45:01
5 Contro. 3 Alarm 01.06.03 17:58:22
4 Contro. 4 Warn. 28.05.03 20:01:22
3 Contro. 5 Warn. 27.05.03 07:58:00
Low level
Help Menu
End
T
set
T