![Lauda RE 1050 G Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/lauda/re-1050-g/re-1050-g_operating-instructions-manual_4088318085.webp)
ECO GOLD
28.04.10/ YACE0088
Maintenance
85
9.3.1
Error store
With the ECO GOLD thermostats there is an error store in which up to 140 errors and alarm messages
can be stored for the analysis and localisation of errors.
−
Access to the menu level is obtained by pressing
.
−
Selection and confirmation of
Æ
Setup
Æ
Device Status
Æ
Errorstore
.
−
The last message is located in the first position.
−
Each message line can be marked with
or
. In the
footer the message appears in plain text.
−
The relevant module which is causing the message is dis-
played under Source.
−
Code is the number to which the message described in
plain text is assigned.
−
Type: Alarm, warning or error.
9.3.2
Device data
The most important device parameters are displayed under the menu point Device data.
−
Access to the menu level is obtained by pressing
.
−
Selection and confirmation of
Æ
Setup
Æ
Device Status
Æ
Device data
.
9.3.3
Software version
Under the menu point SW version the appropriate software versions are displayed, depending on the
device type and connected modules.
−
Access to the menu level is obtained by pressing
.
−
Selection and confirmation of
Æ
Setup
Æ
Device Status
Æ
SW version
.
−
Only the modules which are connected are displayed with
the software version number.
No. Source Code Type
Date
Time
5 Control
29 Error
17.03.10 16:32
4
Safety 3
Alarm
16.03.10
10:32
3 Control
4 Warn
15.03.10 16:41
2 Safety
29 Error
12.03.10 17:02
1 Control
36 Error
11.03.10 08:04
Safety Overtemperature
Display
סּ
OK
Standby
ESC
סּ
OK
Standby
Device data
T_int
22.23°C
T_lp
-86.33°C
T_chip
78.72°C
T_b_s
23.04°C
T_a
-96.33°C
T_max_s
85°C
T_mos
30.93°C
ESC
סּ
OK
Standby
SW version
Control
1.20
Safety
1.20
Cool
1.10
Ext Pt
1.15
Analog
2.10
Serial
2.05
Contact
2.08
Содержание RE 1050 G