Interface modules
28.04.10/ YACE0088
8.6
Commands and error messages applicable to the RS 232/485 interface mod-
ule and to the USB interface
8.6.1
Write commands (data issued to the thermostat)
Command Meaning
OUT_PV_05_XXX.XX
Specify external temperature via interface
OUT_SP_00_XXX.XX
Set-value transfer with max. 3 places before the decimal point and max. 2 places
after it.
OUT_SP_01_XXX
Pump power level 1 to 6
OUT_SP_02_XXX
Cooling operating mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).
OUT_SP_04_XXX
TiH outflow temperature limit, upper value
OUT_SP_05_XXX
TiL outflow temperature limit, lower value
OUT_PAR_00_XXX.X
Setting of the control parameter Xp.
OUT_PAR_01_XXX
Setting of the control parameter Tn (5...180 s; 181 = Off).
OUT_PAR_02_XXX
Setting of the control parameter Tv.
OUT_PAR_03_XXX.X
Setting of the control parameter Td.
OUT_PAR_04_XXX.XX
Setting of the control parameter KpE.
OUT_PAR_05_XXX
Setting of the control parameter TnE (0...998 s; 999 = Off).
OUT_PAR_06_XXX
Setting of the control parameter TvE.
OUT_PAR_07_XXX.X
Setting of the control parameter TdE.
OUT_PAR_09_XXX.X
Setting of the max. outflow temperature limit.
OUT_PAR_10_XXX.X
Setting of the control parameter XpF.
OUT_PAR_11_XXX
Setting of the control parameter TnF (5...180 s; 181 = Off).
OUT_PAR_12_XXX
Setting of the control parameter TvF.
OUT_PAR_13_XXX.X
Setting of the control parameter TdF.
OUT_PAR_14_XXX.X
Setting of the setpoint offset.
OUT_MODE_00_X
Keypad: 0 = released / 1 = locked (corresponds to: "KEY").
OUT_MODE_01_X
Control: 0 = int. / 1 = ext. Pt100 / 2 = ext. Analog / 3 = ext. Serial.
OUT_MODE_03_X
Keypad remote control unit Command: 0 = released / 1 = locked.
OUT_MODE_04_X
Setpoint offset source: 0=normal / 1=ext. Pt / 2=ext. analog / 3=ext. serial.
START
Switches the device on (from Standby)
STOP
Switches the device in Standby (pump, heating, chiller off).
RMP_SELECT_X
Selection of program (1...5) to which further commands are to refer. When the
device is switched on Program 5 is selected.
RMP_START
Start the programmer.
RMP_PAUSE
Stop the programmer.
RMP_CONT
Start the programmer again after a hold.
RMP_STOP
Terminate the program.
RMP_RESET
Delete program (all segments)
RMP_OUT_00_XXX.XX_XXXXX_XXX.XX_
X
Sets programmer segment (temperature, time, tolerance, and pump level). A
segment is appended and assigned appropriate values.
RMP_OUT_02_XXX
Number of program loops: 0 = endless / 1...250.
RMP_OUT_06_XXX.XX
Programmer tolerance setting (0 = off / 0.01 °C...450.00 °C) All the following
segments are assigned this tolerance setting.
−
A " " (space character) is also admissible for "_".
−
Response from thermostat "OK" or with an error " ERR_X" (RS 485 interface
e.g. "A015_OK" or with an error "A015_ERR_X".)