Proline Edition X cooling thermostats
YACE0100 / October 4, 2018
37
7.7
Important settings
7.7.1 Temperature setpoint setting
The setpoint is the temperature, which the thermostat should reach and maintain constant.
Master
(main level)
SEt
Press key until
SEt
(Setpoint) appears.
Press, display flashes.
or
Enter the setpoint with the two keys (
Section 7.4.1 General key
Wait 4 seconds or
Display flashes 4 seconds
new value is automatically accepted,
or
value is accepted immediately with Enter key.
For safety reasons the setpoint can only be set up to 2 °C above
upper limit of the operating temperature range for the relevant
device type.
In the following cases, the manual setpoint entry is blocked:
Setpoint is taken from the analog module, from the programmer in
the Command remote control or via the serial interface.
When the setpoint temperature is to be lowered, it may take up to
one minute before the blue LED
lights.
Содержание Proline RP 1290 C
Страница 2: ......
Страница 4: ......
Страница 115: ...Proline Edition X cooling thermostats YACE0100 October 4 2018 115 Blank page ...
Страница 116: ...Proline Edition X cooling thermostats 116 YACE0100 October 4 2018 Blank page ...
Страница 118: ......
Страница 119: ......