Proline heating thermostat
43 / 123
Command
Control variable
Open the device parameter menu with the soft
key Menu .
With the cursor keys, change further to:
Control
Control variable
.
Intern
is currently active.
Select other control variables (only displayed
when present) with or and confirm with
or End ,
or quit the window with without making any
changes.
7.7.5
Current consumption from the mains
If your mains fuse is rated below 16 A, the current consumption can be reduced in steps from 16 A to 10 A using this
function. Of course the maximum heating power of 3.5 kW is then also reduced accordingly. Take into account wheth-
er other loads are still connected to the fused circuit or whether your Proline Thermostat is the only load.
or
Change temperature probe source:
1
for the internal probe,
EP
only when an external probe is connected,
EA
only when an analog module is connected and configured,
ES
only when a serial module is connected and is continuously receiving ac-
tual values from a PC.
Wait 4 seconds or
Display flashes 4 s
new value is automatically accepted, or
value is immediately accepted with Enter key.
If
EP
EA
or
ES
has been selected,
then the green
LED indicates that the control has regulated to the exter-
nal temperature signal.
Pump Menu
End
T
set
T
fix
Intern
Extern Pt100
Analog module
Digital module
Control variable
12!45
°C
Содержание PROLINE P 5 C
Страница 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Страница 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Страница 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Страница 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Страница 118: ......
Страница 119: ......