Proline Edition X heating thermostats
54
/
10/9/2018
7.8.7 Entering the offset of the internal temperature probe
If, during checking with a calibrated reference thermometer probe, a deviation is found, then the offset
(i.e. the additive part of the characteristic) of the internal measuring chain can be adjusted with the fol-
lowing function. The reference thermometer must be dipped into the bath according to the details on the
calibration certificate.
Master
CAL 1
3x
to
MNEnu
2x
to
CAL..
Call the offset adjustment of the internal temperature probe
CAL..
.
CAL 1
is displayed.
3 sec
The actual value of the bath temperature is displayed flashing.
or
Set the value which you have read from the calibrated reference
measurement device (with glass thermometers consider the correc-
tion where applicable!).
The entered value is accepted and
donE
is displayed.
Command
Calibration
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to:
Settings
Calibration
intern Pt100
Calibration
.
The window shown on the left appears.
Confirm selection with
.
CAL 1
°C
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Содержание PROLINE Edition X P 12 C
Страница 2: ......
Страница 4: ......
Страница 15: ...Proline Edition X heating thermostats YACE0099 10 9 2018 Controls and functional elements 15 Proline P 50 C ...
Страница 109: ...Proline Edition X heating thermostats YACE0099 10 9 2018 Technical data 109 Pump characteristics measured with water ...
Страница 112: ...Proline Edition X heating thermostats 112 YACE0099 10 9 2018 Blank page ...
Страница 114: ......
Страница 115: ......