5 Commissioning
5.1
LAUDA heat transfer liquids
DANGER!
Use of unsuitable heat transfer liquid
Fire
Select a heat transfer liquid with a temperature range suit-
able for the application.
Please note:
If the heat transfer liquid reaches the lower limit of the temperature
range, the temperature control properties can be expected to deterio-
rate as a result of the increase in viscosity. Therefore, only fully utilize
this temperature range when absolutely necessary.
Never use contaminated heat transfer liquids. Contamination in the
pump housing can cause the pump to seize and the device to shut down.
Observe the safety data sheet for the heat transfer liquid. You can
request a copy of the safety data sheets at any time, if necessary.
Table 27: Permitted heat transfer liquids
Designation
Chemical name
Working tem-
perature range in
°C
Viscosity (kin) in
mm²/s (at 20 °C)
Viscosity (kin) in
mm²/s at tem-
perature
Flash point
in °C
Kryo 95
Silicone oil
-95 – 60
1.6
20 at -80 °C
64
Kryo 60
Silicone oil
-60 – 60
3
25 at -60 °C
62
Kryo 51
Silicone oil
-50 – 120
5
34 at -50 °C
120
Kryo 30
Monoethylene glycol/
water mixture
-30 – 90
4
50 at -25 °C
---
Kryo 20
Silicone oil
-20 – 170
11
28 at -20 °C
170
Aqua 90
Decalcified water
5 – 90
1
---
---
Therm 250
Silicone oil
50 – 250
125
25 at 70 °C
300
Therm 180
Silicone oil
0 – 180
23
36 at 0 °C
250
Therm 160
Polyalkylene glycol
and additives
60 – 160
141
28 at 60 °C
260
Table 28: Heat transfer liquid order numbers
Designation
Container size
Catalog number
5 L
10 L
20 L
Kryo 95
LZB 130
LZB 230
LZB 330
Kryo 60
LZB 102
LZB 202
LZB 302
V6
PRO bath thermostats and circulation thermostats
59 / 156
Содержание P 10
Страница 146: ...V6 PRO bath thermostats and circulation thermostats 146 156 ...
Страница 147: ...V6 PRO bath thermostats and circulation thermostats 147 156 ...
Страница 148: ...V6 PRO bath thermostats and circulation thermostats 148 156 ...
Страница 157: ......
Страница 158: ......
Страница 159: ......