Data
Value Unit
Sulfate [SO
4
2-
]
< 70 mg/L
Ratio hydrogen carbonate [HCO
3
-
] / sul-
fate [SO
4
2-
]
> 1 ---
Total water hardness
4.0
-
8.5 °dH
Electrical conductivity
30
-
500 μS/cm
Sulfite (SO
3
2-
)
< 1 mg/L
Free chlorine gas (Cl
2
)
< 0.5 mg/L
Nitrate (NO
3
-
)
< 100 mg/L
Ammonia (NH
3
)
Not permitted ---
Iron (Fe), dissolved
< 0.2 mg/L
Manganese (Mn), dissolved
< 0.05 mg/L
Aluminum (Al), dissolved
< 0.2 mg/L
Free aggressive carbon dioxide (CO
2
)
Not permitted ---
Hydrogen sulfide (H
2
S)
Not permitted ---
Algae growth
Not permitted ---
Suspended matter
Not permitted ---
4.9.2 Connecting the cooling water
This section is relevant for:
Devices with refrigerating machine
Devices with cooling coil
Refer to
Chapter 11.2 “Cooling capacity and cooling water”
for further information about the cooling water pres-
sure, cooling water temperature and diameter of the cooling water
hoses.
Please note:
Connect the cooling water inlet and outlet according to the labeling on
the device. The inlet and outlet of the cooling water supply may not be
interchanged.
The hoses used for the cooling water circuit must be suitable for the
temperature range specified. Also observe the permitted hose diameter.
Secure the hose nozzles or the coupling connectors to the hoses using
hose clips.
Fig. 23: Cooling water input and output
V6
PRO bath thermostats and circulation thermostats
57 / 156