Unit
MC 250
MC 350
MC 600
MC 1200
GWP
(100a)
*
---
1430
1430
1430
1430
CO
2
equivalent
t
0.1
0.1
0.3
0.8
Table 14: Devices with 100 V; 50/60 Hz
Unit
MC 250
MC 350
MC 600
MC 1200
Refrigerant
---
R-134a
R-134a
R-134a
R-134a
Maximum filling weight
kg
0.092
0.095
0.31
0.57
GWP
(100a)
*
---
1430
1430
1430
1430
CO
2
equivalent
t
0.1
0.1
0.4
0.8
Global Warming Potential (GWP), CO
2
comparison = 1.0
* Time frame 100 years - according to IPCC IV
11.4
Hydraulic circuit
MC 250
MC 350
MC 600
MC 1200
Filling volume
L
2 – 4
4 – 7
4 – 8
7 – 14
maximum flow rate
l/min (water 20 °C)
16
16
35
35
Maximum discharge
pressure
bar (water 20 °C)
0.35
0.35
1.30
1.30
Pump connection
(clear Ø in mm)
Hose nozzle
½"
(10)
Hose nozzle
½"
(10)
G
¾
(15), hose
nozzle ¾"
G
¾
(15), hose
nozzle ¾"
Drain tap
Connection
G
½"
G
½"
G
½"
G
½"
Overflow connection
(clear Ø in mm)
Hose nozzle
½"
(10)
Hose nozzle
½"
(10)
Hose nozzle 16
mm (12)
Hose nozzle 16
mm (12)
11.5
Voltage-dependent data
Table 15: Power consumption
MC 250
MC 350
MC 600
MC 1200
kW
kW
kW
kW
230 V; 50 Hz
0.23
0.50
0.70
1.15
220 V; 60 Hz
0.23
0.50
0.70
1.15
115 V; 60 Hz
0.23
0.50
0.75
1.10
100 V; 50/60 Hz
0.23
0.50
0.75
1.10
V07
Microcool