6
Operation
6.1
Switching on the device
WARNING!
Overheating of the pump
Device damage
Never operate device without heat transfer liquid.
Personnel:
Operating personnel
1.
Switch on the device using the mains power switch.
A signal tone sounds. The actual temperature is shown
on the display. The pump is started. The refrigeration unit
is started after approx. 2 minutes. The blue LED lights if
the refrigeration unit is active.
2.
Depending on the size of the consumer, heat transfer liquid
must be refilled if necessary. Monitor the level indicator for
this.
Further information about refilling heat transfer
heat transfer liquid’ on page 35
.
6.2
Default display and menu items
1.
Press the ENTER button to reach the menu items from the
default display of the actual temperature.
If no button has been pressed for longer than 4
seconds, you exit from the menu item or input
window.
2.
Scroll from menu item to menu item using the arrow buttons.
3.
Press the ENTER button at the selected menu item.
The display flashes.
4.
The value or the setting can be changed using the arrow but-
tons.
5.
The changed value or setting is applied immediately by
pressing the ENTER button.
If no button has been pressed for longer than 4
seconds, changed values or settings are applied
automatically and you exit from the menu item or
input window.
Operation
Microcool
38
Содержание MC 1200