1.
In the
Control parameter
menu, tap the [Parameter sets] button.
A list of available parameter sets is displayed.
2.
Tap the required parameter set.
The selected parameter set is displayed with a light background.
3.
Tap one of the following buttons to use the selected parameter set:
[Display]: Displays the control parameters saved in the parameter
set. You can read the data but you cannot modify it.
[Load]: Loads and activates the parameter set. The previously
valid control parameters are overwritten.
[Delete]: Removes the parameter set when you confirm this with
the [OK] button.
7.4 Ramp function
The ramp function allows you to conveniently enter temperature changes
that occur over any period of time. The function is particularly useful for
recording extremely small temperature changes (for example, 0.1 °C/day).
If the current bath or outflow temperature (for example, 142.4 °C) should
decrease by 80 °C within 5 days, a temperature change of 80 °C must be
entered. A time of 5 days must also be entered.
Example:
The ramp function remains active until it is stopped manually or the
temperature limit values T
ih
or T
il
are reached.
1.
Open the Home window.
2.
Touch the [Ramp] tile.
The Ramp function submenu appears.
3.
Touch the [Temperature increase] button.
Enter the temperature change in the input window (not the end
temperature!). You can also enter negative temperature changes.
4.
Press the [OK] button to return to the previous screen with the new
setting configured.
5.
Move the slide switch on the [Time unit] button to the required time
unit.
ss - Seconds
mm - Minutes
hh - Hours
dd - Days
6.
Touch the [Duration of temperature increase] button.
The Input window opens.
7.
Enter the numerical value in the input window (unit not available!).
8.
Press the [OK] button to return to the previous screen with the new
setting configured.
Displaying and loading parameter sets
V1
Command Touch operating unit
35 / 51