Integral XT
YAWE0028 / 24/06/2016
Starting up
109
7.16.5 Pump-motor supervision: Overload or blockage
%
‰‰
The SelfCheck Assistant monitors the Vario Pump:
1. Alarm sounds as dual-tone signal for pump-motor overload or
blockage.
2. Display of
bL0C
signals blockage.
3. The red LED
above the fault triangle
flashes.
à
Heater switches off on both poles;
à
Pump and refrigerating unit are switched off electronically.
−
Rectify cause of fault (
þ
9.4).
−
Press the enter key.
−
Also press this key if the unit has been switched off in the fault
state.
Command
Pump-motor alarm!
−
Pump-motor alarm! is shown in the display and signifies that unlocking is
only possible on the Master control panel.
7.16.6 Pump-motor supervision: Dry running
%
‰‰
The SelfCheck Assistant monitors the Vario Pump:
1. Alarm sounds as dual-tone signal when the pump runs without
liquid.
2. The display of
PuLEU
signals that the SelfCheck Assistant
has detected a pump low level.
3. The red LED
above the fault triangle
flashes.
à
Heater switches off on both poles;
à
Pump and refrigerating unit are switched off electronically.
+
The cause of the failure of the level measurement with the floatation sensor must be
found and rectified (
þ
9.4).
−
Press the Enter key.
−
Also press this key if the unit has been switched off in the fault
state.
Command
Alarm! Low level (pump)
−
Alarm! Low level (pump) is shown in the display and signifies that
unlocking is only possible on the Master control panel.
Pump alarm
bL0C
°C
Pump alarm
PuLEU
°C