6.4.5 Adapting control parameters
Personnel:
n
Specialized personnel
Press the
[ESC]
soft key to return to the previous display without
making any changes.
1.
Press the Enter key to open the menu.
2.
Select the menu items
Setup
Control
Control parameter
.
If the control variable extern is active, the external control param-
eter is displayed on the display.
If the control variable internal is active, the internal control
parameter is displayed on the display.
3.
Scroll to another control parameter and select it with the Enter key.
An entry window opens. You can now change the numerical value.
The displayed values
Max:
and
Min:
specify the limits for the
value entry.
4.
Press the [OK] to confirm the new value.
The new value is active.
Control parameter Tv
n
With
Tv manual/auto
you can define whether the control parameter
Tv
and
Td
or
Tve
,
Tde
and
Prop_E
are set manually or automatically. If the
automatic setting is enabled, these control parameters are displayed with
a lock and cannot be selected.
To be able to set these control parameters manually, change the control
parameter
Tv manual/auto
to manual setting.
6.5 Self adaptation
The self adaptation function can be used to automatically find suitable con-
trol parameters for the respective application. Self adaptation can only be
executed on a device with active cooling.
The self adaptation function performs a test run of the constant temperature
equipment to determine the parameters. The device and application must be
ready for operation.
For Integral XT, self adaptation is carried out with the currently set pump
level. The best results can be expected at the highest pump levels. The
pressure control of the pump can be used, but will worsen the result. If
the current consumption of the device is significantly limited, the result
deteriorates and the duration of the self adaptation increases.
The test run must be performed on a passive application. In other words,
changes cannot be made to the application and exothermic or endothermic
reactions may not occur during the test run.
The test run takes between 30 minutes and a maximum of 3 hours,
depending on the application. During the run, the temperature of the heat
transfer liquid deviates a maximum of ±15 Kelvin from the defined setpoint.
After completion of the test run, the newly determined values of the control
parameters are accepted and stored in the [Control parameter] menu. The
old control parameters are overwritten.
Fig. 53: Change control parameters
V4
Integral process thermostats with flow control unit
90 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...