Required equipment for descaling with a pump (drum pump):
n
a container with approx 20 liter volume
n
a pump
n
Hoses between container and pump and between pump and cooling
water circuit
n
Hose between cooling water circuit and container
Required equipment for descaling with a funnel:
n
Two containers with 10 to 20 liter volume
n
A funnel
n
Hose between funnel and cooling water circuit Place the funnel as high
as possible, so that the device fills with descaler quickly.
n
Hose between cooling water circuit and container
Protective equipment:
n
Safety glasses
n
Protective gloves
n
Protective work clothing
NOTICE!
Risk of refrigerant escaping from cooling water circuit
Device damage
Before performing cleaning work and decommissioning the
device, or if there is a risk of freezing, drain the cooling
water circuit of the cooling unit using compressed air or
an industrial vacuum cleaner (watertight). To avoid dam-
aging the valve membrane during cleaning, only blow com-
pressed air into the inflow connector while using an indus-
trial vacuum cleaner to vacuum only from the outflow
connector.
1.
Switch the device into standby.
2.
Press the Enter key to open the menu.
3.
Select the menu items
Fill mode
Water valve
open
on the
operating unit.
The water valve opens.
4.
Fill the device with descaler-water mixture via the hose on the cooling
water inlet.
5.
Recirculate the descaler/water mixture or refill continuously.
6.
Leave the descaler-water mixture to work (see table below).
7.
Flush the device thoroughly with clean water (see table below).
8.
Drain cooling water circuit Blow compressed air through the return
connector of the cooling water circuit.
Descaling the cooling water circuit
Fig. 81: Descaling
V4
Integral process thermostats with flow control unit
129 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...