3.5 Mains switch
The devices have a mains switch on the right side of the device. This is
designed either as a toggle switch or as a rotary switch. Position [0] switches
the device off, position [I] switches it on.
Mains switch on the single-phase device
The toggle switch is also designed as a safety switch. If the current is
too high, the toggle switch trips and disconnects the device from the
mains supply. The device can be used again by switching the toggle
switch to position
[I]
. If the safety switch trips again, contact the
LAUDA Service department.
Mains switch on the three-phase device
The automatic fuse may only be switched on by a qualified electri-
cian!
The rotary switch is not designed as a safety switch. The three-
phase device has a separate automatic fuse built in. If the automatic
fuse trips, contact the LAUDA Service department.
3.6 Power supply at IN 2560 XTW and IN 2560 PW
NOTICE!
Use of a mains cable with impermissible dimensions
Device damage
The specifications on the type plate of the device are deci-
sive for the cross-section of the mains cable.
Comply with the national regulations of the respective
country when installing the mains cable.
Notes for electric installation on site:
n
The mains cable must be installed by a qualified electrician.
n
The mains cable must be installed on site by the user. The installation is
implemented with hard wiring or with an approved plug with an adequate
rating.
n
For three-phase devices the rating of the circuit breaker must match
the power consumption loading of the device. The value will be found
on the type plate. In each case, select the next higher value. Using an
excessively high rated circuit breaker is not permissible.
Installation of the mains cable on site
V6
Integral Process Thermostats and High-Temperature Thermostats
25 / 198
Содержание IN 1030 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198 ...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198 ...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198 ...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198 ...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198 ...
Страница 199: ......