ECO Gold
51 / 137
7.3.4
Graphics window
The ECO GOLD thermostats offer you the possibility of displaying temperature traces graphically.
In the graphics window the following information is displayed depending on
the setting:
T
set
set-point temperature (grey)
T
int
internal bath temperature (green)
T
ext
Temperature on the external consumer, external
temperature sensor (blue).
7.4
Basic setup
In this section the settings required for using the device as prescribed are summarized. For more extensive settings refer
to the appendix.
7.4.1
Setting the overtemperature switch-off point T
max
Warning!
Overheating due to entering an incorrect T
max
and set-point tempera-
ture
Burns, scalds, fire
Set T
max
in each case according to the heat transfer liquid
used T
max
must be below the flash point.
-
Hold the key
pressed during the complete setting procedure:
-
Press the enter key .
The entry window appears. The cursor flashes under the T
max
value. The
maximum and minimum adjustable temperature values are displayed.
-
Change the value with or .
Note: With a longer depression the figures increment faster.
-
Single figures can be selected by pressing or .
-
Confirm your choice with the enter key .
On releasing
you are returned to the menu level without any change.
For T
max
the following applies: 5 Kelvin above the required maximum bath temperature, but below the flash point of the
heat transfer liquid.
Содержание ECO GOLD
Страница 1: ...Operation manual ECO Gold ...
Страница 2: ......
Страница 4: ......
Страница 81: ...10 2019 i ECO Gold 81 137 W Warnings 55 92 X Xp 106 Xpf 108 ...
Страница 82: ...82 137 ECO Gold 10 2019 i ...
Страница 83: ...10 2019 i ECO Gold 83 137 ...
Страница 84: ...84 137 ECO Gold 10 2019 i ...
Страница 85: ...10 2019 i ECO Gold 85 137 Appendix with settings ...
Страница 121: ...10 2019 i ECO Gold 121 137 Click on the tab with the mouse and then on the Device manager ...
Страница 130: ......
Страница 132: ......
Страница 133: ......
Страница 134: ......
Страница 135: ......