ECO GOLD
25/08/2011/ YACE0088
Appendix with settings
85
A.8
Restoring the factory setting of the internal temperature sensor
If the offset has been adjusted, the factory setting can be restored again.
−
Access to the main menu level is obtained by pressing the
enter key
.
−
Selection and confirmation of
Æ
Setup
Æ
Calibration
.
The adjacent menu window appears.
−
Select
and
confirm
Factory calibration
with or and
.
The adjacent menu window appears.
−
Select
"yes"
using or and confirm with
to restore
the factory settings.
−
By pressing "no" (ESC) or you are returned to the menu level without any change.
ESC
סּ
OK
Standby
Calibration
Calibration
Factory Calibration
.
ESC
סּ
OK
Standby
Factory calibration
no
yes
.