SPA Spectrum Analyzer 

internal battery

The internal battery is a high capacity Lithium-ion polymer 1000mAh. This is the same battery technology 
used in cell phones and modern laptops and tablet computers. 

The power switch is a 

true hard switch

 connecting the battery in ON position, fully isolating the battery in the 

OFF position. As the battery is fully disconnected in the OFF position, you can keep your RF 

Spectrum 

analyzer

 stored for months and the battery will keep the charge intact, as opposed to cell phones or laptops 

which typically implement a 

soft switch

 which slowly discharge the battery by drawing a microamperes 

continuously from it. This doesn't happen in your 

LATNEX hand-held RF analyzer device

Note there is actually a self-discharge effect in the battery but it is so negligible that you can store your 
device for a year or more and a battery will not be depleted. 

Charging the internal battery 

To charge the internal battery, plug it in to a powered USB port or USB wall-wart charger via the USB mini 
socket. For the internal charger to access the battery, you must set the power switch to ON position, 
otherwise the device will be running from USB but will not use or charge the internal battery. 

The RF signal analyzer 

may draw up to 500mA according to USB standard, and may take up to 3hs max to 

charge a fully discharged battery in normal conditions. We recommend the first time to be charged for up to 
8hs. 

IMPORTANT 

Never charge your device unattended or in any place where there is a risk of fire. 

Never store 

the SPA Spectrum Analyzer

 in a place where temperature may go

 

beyond 50ºC (122F). 

A car can heat up incredibly quickly in the 

sun

, especially in the summer, and 

damage or reduce the lifetime of your battery if stored in such an environment. 

If your battery doesn't seem to hold a charge or work properly, or if the unit gets 

hot when charging, switch the power button OFF immediately and review and/or 

replace your battery. 

Never use 

any of the LATNEX Spectrum Analyzer

 

models 

with a damaged battery or 

with a Lithium Ion battery different than the one officially supplied by technical 

service, which includes a 

protection circuit for safety. Contact 

EMR Shielding Solutions

 

if you have any 

question on

 

any matter related with your battery or if you need a replacement. 

www.latnex.com/spa 

www.emrss.com/latnex-analyzers

Toll-free number  1(855)267-2582 

[email protected]

17

Содержание SPA-50K

Страница 1: ...User Manual SPA 50K 50KHz 960MHz SPA 3G 15MHz 2700 MHz SPA 6G 15MHz 2700MHz and 4850MHz 6100MHz www latnex com spa www emrss com latnex analyzers Toll free number 1 855 267 2582 info latnex com 1...

Страница 2: ...ndling with SET button and a silicone rubber protector The unit also includes one RF connector 50 ohm impedance standard SMA format Important The internal battery power switch must be in the ON positi...

Страница 3: ...m to 110 dBm Configuration settings such as amplitude reference and frequency span are preserved between sessions stored in internal FLASH memory The main screen automatically displays a small marker...

Страница 4: ...mode in the frequency menu Please check the section Frequency Menu in page 15 DSP mode this indicator shows the actual value being used by the Analyzer The DSP Auto is the recommended setting in the...

Страница 5: ...another You can exit from a menu anytime by using the Return key The first time you click on Menu button Frequency Menu will open This may change as we will see later so every time you click on Menu f...

Страница 6: ...ons enumerates the installed RF modules and how each one is connected to each SMA RF port This example below shows a standard SPA 3G with a 240 960MHz connection on the left port and 15 2700MHz on the...

Страница 7: ...y useful to select frequency range between two frequencies of interest with no need to manually calculate the Center or the Span The valid values for Frequency and Span are defined based on LATNEX SPA...

Страница 8: ...ctrum Analyzer main screen Note While SPA Spectrum Analyzer screen is enabled the Left and Right buttons can be used to increase or decrease the Start Stop frequency virtually moving the visible spect...

Страница 9: ...a 10MHz span each of the 112 measurement steps represents 10MHz 112 0 089MHz or in other words 89KHz2 The range of frequencies represented on each sweep step is a level of uncertainty screen resoluti...

Страница 10: ...ortunately this is not the case with the RF analyzer The design of the internal DSP includes sophisticated internal tables and algorithms to make the best possible choice at all times automatically No...

Страница 11: ...e anytime you want to detect activity for a long period of time It is a very powerful mode to detect activity in busy bands such as 2 4 GHz As this is a cumulative mode while on Spectrum Analyzer scre...

Страница 12: ...or 30dB nominal wideband attenuator use for linear measurement up to 0dBm LNA low noise wideband preamplifier use for linear measurements up to 40dBm Note this option requires more battery power than...

Страница 13: ...on screen by using Left and Right buttons When this mode is selected the standard functions to move Center Frequency or Span are disabled o None The marker is not displayed DSP The DSP Digital Signal...

Страница 14: ...reset with data stored is selected the last line of the screen shows the Start Stop frequency range of the preset to help on determining the specific preset configuration before using it This line wil...

Страница 15: ...een shows when additional text is available for scroll Use UP and DOWN arrow key buttons to scroll text one line at a time or LEFT and RIGHT arrow key buttons to scroll text full page at a time Help o...

Страница 16: ...tch the transmitter ON without a proper antenna you may damage the RF module of your SPA Spectrum Analyzer as the power amplifier will not have a load to feed Think of the equivalent to an audio ampli...

Страница 17: ...r switch to ON position otherwise the device will be running from USB but will not use or charge the internal battery The RF signal analyzer may draw up to 500mA according to USB standard and may take...

Страница 18: ...power switch is in the ON position the battery will charge However you can get additional information when using the Battery specific menu Inside this INTERNAL BATTERY charger screen it is easier to r...

Страница 19: ...voltage the internal charger considers 100 capacity This reference is subject to change over time as the internal battery reduces total capacity over many charge cycles As a reference 500 charge cycle...

Страница 20: ...ernal battery charger will take care of the battery conditions once it gets powered back from the USB and will charge the battery using an internal efficient algorithm For that to work properly when t...

Отзывы: