background image

ventus

 with smd12

ventus

Operating 

Manual

R-0162

Ver. 2.0

Содержание Ventus with smd12

Страница 1: ...ventus with smd12 ventus Operating Manual R 0162 Ver 2 0 ...

Страница 2: ...n 4 Section 1 1 Operational Requirement Electrical Specification 4 Section 1 2 Operational Requirement Environment 4 Section 1 3 Storage Requirement Environment 4 Section 2 Laser Safety 5 Section 3 System Specifications 6 Section 3 1 Parts List 6 Section 3 2 Weights and Dimensions 7 Section 4 PSU Configuration Drawings 8 Section 5 Laser Operation 9 Section 5 1 Switching the Laser on 9 Section 5 2 ...

Страница 3: ...mal Operating Temperature Range Laser Head 22ºC to 37ºC Maximum Operating Temperature Laser Head 40ºC Maximum Operating Temperature PSU 60ºC For optimal performance the laser head should be mounted onto an appropriate heatsink in a stable environmental temperature The heatsink requirements will depend on the ambient temperature of the operating environment and the operating power of the laser sect...

Страница 4: ...ending on the purchase specification some or all of the following items may also be included External Power Supply Source Mains AC to 12V DC Desktop Module IEC Mains Lead Interlock Dongle to enable the laser system Yellow spot on casing with Yellow LED Breakout Cable to enable RS232 connection RemoteApp software suite Operating Manual delivered on a USB flash drive Note If a Blue Red spot Interloc...

Страница 5: ...d this can be achieved using the supplied Green spot Interlock Dongle 5 Depending on the start up conditions set for the system the laser may now start to emit laser light Caution Care must be taken to prevent direct exposure to the laser beam before the system is enabled Turn page Turn page Fig 4 1 Fig 4 2 LED Functions 1 Red Laser Enabled 2 Red Power Fail 3 Red Over Temp Head 4 Red Over Temp PSU...

Страница 6: ...s can be found in Section 5 5 The remote Interlock Enable and Enable LED can be wired in accordance with the diagrams below Fig 5 4 Fig 5 5 to be used as part of a laboratory interlock safety circuit If either Interlock is broken the system will shut down Connections to the control port shall be potential free i e isolated from mains voltage by a barrier rated at 3KV e g Double Insulation Etc The ...

Страница 7: ...tal time the system has been powered Laser Enabled Time Total time the diodes have been powered Laser Operation Time Total time the diodes have been powered 1A Turn page Turn page Section 5 5 RS232 Port Functionality Control of the laser can be achieved via the RS232 port using a terminal emulator such as HyperTerminal or PuTTY This allows the operator to Turn the laser on off Control the laser po...

Страница 8: ...of up to 5 as a result of power meter variation Section 6 Cooling Requirements and Power Consumption Section 6 1 Cooling Requirements The laser has a characteristic warm up period before it reaches specification this time depends partly on the heat sink to which it is attached However the typical warm up time is 10 minutes from switch on In order for the laser to perform to specification it must b...

Страница 9: ...f the laser window may be necessary To perform this procedure the laser must be turned off and using optical cloth dampened with research grade methanol the laser window must be wiped gently Always follow the instructions given in this Operating Manual Never touch the connector on the laser head with anything other than the Umbilical cable provided and always follow the connection instructions in ...

Страница 10: ...on under this warranty is limited to the replacement or repair of the product which having been returned to the factory is found to be defective and where the defect was not caused by factors external to the product Any replacement part product is under warranty for the remainder of the initial product warranty period Section 10 Support Should the laser fall below acceptable specification performa...