![Larson Davis Spartan 730 Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/larson-davis/spartan-730/spartan-730_reference-manual_3719581084.webp)
Spartan 730/730IS Dosimeter Manual
A-14
Spartan Model 730IS also complies with the following hazardous environment standards
:
A.2.2 Compliance Testing Notes
For compliance testing, an electrical signal is applied to the dosimeter using the ADP106 “ADAPTER INPUT BNC TO 1/4” MIC 12
pF”. Electrical tests are performed in a stable temperature environment that does not vary more than ±2° C during the test. The
dosimeter is allowed to stabilize to the ambient temperature prior to testing.
Reference Orientation
The unit is angled at about 45 degrees so that the microphone is pointing up. The unit is facing toward the radio-frequency
electromagnetic field emitter or receiver.
Normal Mode of Operation during Tests
The Spartan 730 is running a measurement while set up in the reference orientation with time history enabled.
Magnetic Field
The sound exposure level of the Spartan 730 was always within ± 0.1 dB of the original level while exposed to a uniform
magnetic field strength for 1 hour. The unit was set up in the reference orientation.
Radio-Frequency Electromagnetic Fields
Operating the Spartan 730 in proximity to a source of electromagnetic radiation equal to 10 V/m may cause up to ± 1.5 dB
variation on the sound exposure level reading. The Spartan 730 meets all of the standards of section 15 in IEC 61252 (2017) at a
sound level of 85 dB. The mode of operation of the Spartan 730 that produces the least immunity to the effects of power and
radio frequency fields is with the Spartan 730 running set up in the reference orientation with the exception that the unit is
facing perpendicular from the electromagnetic field emitter (90 degrees or 270 degrees).
Electrostatic Discharge
Operating the Spartan 730 in proximity to a source of electrostatic discharge causes no discernible variation or difference in
the functionality of the dosimeter.
Mechanical Vibration
No limitations occur on the Spartan 730 operation resulting from mechanical vibrations.
Radio Frequency Emission
The mode of operation of the Spartan 730 that produces the greatest radio frequency emission levels is with the Spartan 730
set to run with the LCD light on being charged by an external wireless charger. The orientation of the unit does not make a
difference.
A.2.3 ANSI S1.25 Compliance
I.S. (ATEX, IECEx)
EN IEC 60079-0:2018, EN IEC 60079-11:2012