Spartan 730/730IS Dosimeter Manual
D-4
C.3 Intrinsic Safety Ratings and Information
Special Conditions for Safe Use
E
quipment has been assessed for connection to Um: 5.5VDC. This voltage
shall be maintained in accordance with the guidance given with IEC 60079-14.
Other Safety Information
•
Charge ONLY outside hazardous areas using the appropriate charger
or
other equipment approved by Larson Davis.
•
Charge the device at temperatures between 10
°
C and 40
°
C in indoor
conditions.
•
Do not attempt to physically connect the device to other equipment
within ex-hazardous areas.
•
Do not expose the device to acid or alkaline substances, or solvent-based
cleaner.
•
Calibrate the microphone outside ex-hazardous areas only.
•
Do not replace the battery yourself or attempt to install a lithium battery.
Without exception, contact Larson Davis for authorized battery
replacement.
•
Only use accessories approved by Larson Davis with this device.
Applicable Standards
I.S Ratings/Markings
Serial Numbers
IEC 60079-0: Edition 7; EN 60079-0:2018
IEC 60079-11: Edition 6; EN 60079-11:2012
UL 913: Edition 8
UL 60079-0: Edition 7
UL 60079-11: Edition 11
CSA C22.2 No. 60079-0: 19;
CSA C22.2 No. 60079-11: 14
UL 61010-1: Edition 3 + R:29Apr2016 CAN/CSA C22.2 No. 61010-1-2012 + U1
ATEX/IECEx
II 1G Ex ia IIC T165 ºC Ga
I 1M Ex ia I Ma
IECEx ETL 19.0045X ITS20ATEX205538X
-10 ºC ≤ TAMB ≤ +50 ºC
CAN/USA
Class I, Zone 0, AEx ia IIC T165 ºC Ga
Class I, Division 1, Groups A, B, C & D, T165 ºC
-10 ºC ≤ TAMB ≤ +50 ºC
For each serial number YYXXXX, YY is the 2 digit year of manufacture,
and the XXXX is the device sequence number.
I730.01-IS-SD (A) Do not change drawing without approval from the notified body and MSHA.
Any changes could result in an unsafe condition.