Performing Calibration Checks
3-11
TAKE NOTE
Answer Yes to save
the Standard Cal. Level value.
This should be done
immediately following the first
calibration check. Otherwise,
answer No to simply record the
level (and spectrum) in the
Calibration Check History for
trend determination.
FIGURE 3-14
Save New Standard Cal. Level
3.4.2 Automatic Calibration Checks
Calibration checks can also be scheduled to occur automatically on the
Control tab by following these steps:
LEARN MORE
For more
information on setting up
automatic calibration checks,
refer to the SoundAdvisor
Model 831C Sound Level Meter
Manual.
Step 1
Set the run mode to Continuous.
Step 2
Select and enable the Cal. Check option.
Step 3
Set the time of day when the check is to be performed. The
default time is 2:30 AM.
FIGURE 3-15
Scheduled Calibration Checks