Percep�ve Motor Control
Instructions for Use
REF 13501
Lares Research
295 Lockheed Ave
Chico, CA 95973
Device Specifica�ons
⚠
Caution: This device is not designed
for use in an explosive atmosphere
(anesthetic gas).
Contents:
The Percep�ve Motor Control System
consists of:
• Motor hose output
• A pneuma�c 4-hole connec�on input
• An electrical power supply and plug cord
• Percep�ve motor control box
• Tablet
Iden�fica�on:
Electronically controlled unit for den�stry
allowing opera�on of the Percep�ve Motor
with variable speed using the dental unit
rheostat pedal in conjunc�on with the
included tablet.
Intended Use:
• Product intended for professional use
only. Use in den�stry for general den�stry
work. The system is designed to control a
Percep�ve Motor which can drive a dental
handpiece (gear ra�o 1:1, 10:1, or 1:5)
fi�ed with appropriate burs.
• Any use other than that for which this
device is intended is prohibited and may
prove dangerous.
Classifica�on:
Class IIa in accordance with European
Direc�ve 93/42/EEC concerning medical
devices. This medical device is in
compliance with the legisla�on in force.
Electrical Safety:
According to IEC 60601-1 standard
(General safety for medical Electrical
Equipment), the device shall be classified
as a class II type B device.
The following requirements as specified in
IEC 60601-1 apply:
• Protec�on against electrical shock.
• Ingress of liquids.
• Protec�on against excessive
temperatures and other safety hazards.
Electric insula�on class:
Class 1 per IEC60601-1 (apparatus
protected against electric shock).
Degree of protec�on:
IP40 (protec�on against inser�on of
objects larger than 1mm).
RX Only
70525.AW REV B 11/22
Page 1
Electromagne�c Compa�bility:
Electro-medical equipment needs special
precau�ons regarding EMC and needs to
be installed and put into service according
to the EMC informa�on provided in this
document. This unit corresponds to the
electromagne�c compa�bility in
accordance with IEC 60601-1-2 and IEC
80601-2-60 and declara�on by the
manufacturer regarding electromagne�c
compa�bility.
Motor Type:
Brushless type, 3 phase motor,
synchronous with permanent magnets.
Electrical and Pressure Data:
Voltage: 100 – 240 VAC.
Frequency: 50-60 Hz.
Nominal power: 65 W.
Maximum input power: 150W.
Maximum air pressure: 72.5 psi (5 bar).
Minimum air pressure: 43.5 psi (3 bar).
Air consump�on: Less than 40 nl/min
⚠
Caution
If the input air pressure is below the
minimum threshold (17.5 psi), the motor
will not turn on. If operating in variable
speed mode, a minimum of 43.5 psi (3 bar)
is required to meet maximum motor
speed.
Cooling:
Cooling is conducted through compressed
air from the unit.
Spray Air and Water:
Set spray air and water pressure per
contra-angle attachment instructions, but
do not exceed 40 psi (2.8 bar).
Dimensions:
Perceptive Motor Control box; 103 x 84 x
28 mm.
Motor hose length; 1.6M.
Power supply; 178 x 64 x 41mm.
Lenovo M7 Tablet; 176 x 103 x 9mm.
Weight:
Perceptive Motor Control box; 235g (8.29
oz.) without the cable.
Power Supply; 870g (w/ cable).
Tablet; 235g.
Motor Rotation Speed:
From 1,000 rpm to a maximum 40,000 rpm
Motor Direction:
Clockwise and anti-clockwise
Installa�on
Before installing, please read carefully.
Tablet Setup:
Important! Set up the tablet before
connecting it to the Perceptive system.
1. Turn on the tablet by pressing and
holding the power button until the Lenovo
logo appears. (Figure 7)
2. Follow the on-screen prompts to
configure the tablet. You must correctly
connect to WiFi and sign in to a Google
account before downloading the
Perceptive app. Refer to the Lenovo user
guide for appropriate use of the Lenovo
tablet.
3. Open the Google Play store app.
4. Search for the Perceptive app.
5. Install the Perceptive app designed by
Lares Research.
System Installa�on:
1. Chose a loca�on to mount the
Percep�ve motor control box on a flat
surface capable of bearing its weight. It
may be posi�oned horizontally or ver�cally
on, or under, a table, dental unit, or any
other surface but under no circumstances
on the floor. It should be located close to
where the handpiece supply hoses exit on
the delivery system. It is not designed to be
placed on wet surfaces or to come in
contact with liquids.
2. For under-moun�ng clean moun�ng
surfaces with included alcohol wipes. Wipe
away excess alcohol with the provided dry
wipe.
3. Apply two adhesive backed Velcro pads
to one side of the control box. (Figure 1)
A�ach the other Velcro pads to the first
pads by velcro. Remove the adhesive
covers and push firmly against surface to
mount box. Note: if the control box will not
be mounted with the Velcro pads the four
rubber feet may be affixed to one side of
the control box.
4. Connect the power cord to the power
supply (Figure 2). Place the power supply
on the floor or in the electrical outlet box
and plug the power cord into an electrical
outlet.
⚠
Caution
Do not connect control box to a power
supply other than the included unit.
5. Connect the power supply cable to the
motor control box (Figure 3).
6. Connect the Percep�ve motor to the end
of the control tubing; align the connector
pins on the motor with the holes on the
tubing, �ghten clockwise (Figure 4).
7. Connect a 4 hole handpiece hose from
your delivery system to the Percep�ve
motor control and �ghten clockwise (Figure
5). Place the Percep�ve motor in the
vacated handpiece hose nest.
8. Plug the tablet power chord into the
motor control box (Figure 6).
9. Make sure the tablet is turned on. If not,
turn on the tablet by holding the power
bu�on down un�l the Lenovo logo appears
(Figure 7).
10. Plug the tablet power cord into the
tablet (Figure 8) and open the Percep�ve
app. When prompted, allow access to the
microphone to enable voice control.
11. Chose a loca�on to mount the tablet on
a flat surface capable of bearing the weight
of the tablet.(Figure 9) It may be mounted
on a dental unit arm or any other available
surface.
It should be located where the
user can easily view and reach it during
opera�on.
The tablet mount uses a combina�on of
adhesive and suc�on to adhere to its
moun�ng surface
.
Before applying the
adhesive surface, clean the moun�ng
surface with an alcohol pad, then dry with
the dry wipe.
Remove the adhesive cover and depress
the mount’s adhesive surface in the
desired loca�on.
Then depress the lever on
the base of the mount to ac�vate the
suc�on.
12. A�ach the tablet to the tablet mount
(figure 10).
The tablet is retained in the
tablet mount by magne�sm.
13. Secure any loose cords with the
included adhesive “C”clamps.
The Percep�ve Motor Control System is
now ready to use!
Device Opera�on
⚠
Important
• The device must not be started without a
bur inserted into the chuck.
• To ensure that your Contra-angle/micro
motor unit functions with maximum
efficiency, it must be cooled by an air
supply of 5-10 Nl/min on the nose of the
micro-motor.
• Never mount an instrument on a rotating
motor.
• Ensure that the Perceptive Motor hose is
not bent or pinched.
Opera�ng Environment:
Temperature: +10°C (50°F) to +25°C (77°F).
Rela�ve Humidity: 20% and 90%.
Atmospheric pressure: 700 hPa to 1060
hPa. Al�tude: 0 to 3048m (0 to 10,000 �).
Selec�ng A�achment Gear Ra�o:
Use the tablet touch screen (figure 11) to
select from the following available
a�achment gear ra�os (3):
-Green, 10:1 handpiece: 100 to 4,000 rpm.
-Blue, 1:1 handpiece: 1000 to 40,000 rpm.
-Red, 1:5 handpiece: 5000 to 200,000 rpm.
⚠
Caution
Verify that the handpiece gear ratio
corresponds to the ratio selected on the
tablet. The buttons are color coded to
match the handpiece color ring with the
correct ratio.
Any of the attachment gear ratio buttons
can be changed to a different ratio by long
pressing on the button and selecting a
new one.
Setting Maximum Motor Speed:
To set your maximum rpm for the current
gear ratio. You have three options. 1. Select
one of the preset rpm buttons. 2. Slide the
maximum rpm slider bar to the desired
maximum speed. 3. Tap on the RPM read
out, then key in the RPM you want. Click
the blue checkmark to accept the value.
(Figure 11).
The maximum rpm chosen will
display above the slider bar (1).
The factory preset maximum rpm values
for each gear ratio can be changed by
setting the speed you want on the slider
bar, long pressing one of the preset
buttons, and clicking SAVE. Long press the
preset button and hit RESET to go back to
the factory default values.
Motor Rotation Direction:
Press the FWD/REV button (4) on the tablet
to toggle direction (figure 11). FWD is
clockwise and REV is counter-clockwise
while viewing from the back of the
handpiece. (Clockwise is normal status).