7: Line and Tunnel Settings
EDS-MD® Medical Device Server User Guide
49
To Configure Tunnel Serial Settings
Using Web Manager
To configure the Serial Settings for a specific tunnel, click
Tunnel
in the menu and select
Tunnel 1 -> Serial Settings
.
Using the CLI
To enter Tunnel 1 command level:
enable -> tunnel 1 -> serial
Using XML
Include in your file:
<configgroup name=”tunnel serial” instance=”1”>
Packing Mode
With Packing, data from the serial Line is not sent over the network immediately. Instead, data is
queued and sent in segments, when either the timeout or byte threshold is reached. Packing
applies to both Accept and Connect Modes.
Table 7-4 Tunnel Packing Mode Settings
DTR
Select the conditions under which the Data Terminal Ready (DTR) control
signal on the serial line is asserted. Choices are:
Unasserted
TruPort
= the DTR is asserted whenever either a connect or an accept
mode tunnel connection is active with the Telnet Protocol RFC2217 saying
that the remote DSR is asserted.
Asserted while connected
= the DTR is asserted whenever either a
connect or an accept mode tunnel connection is active.
Continuously asserted
Tunnel Packing
Mode Settings
Description
Mode
Configure the Tunnel Packing Mode. Choices are:
Disable
= Data not packed.
Timeout
= data sent after timeout occurs.
Send Character
= data sent when the Send Character is read on the Serial Line.
Threshold
Set the threshold (byte count). If the received serial data reaches this threshold, then
the data will be sent on the network. Valid range is 100 to 1450 bytes. Default is 512.
Timeout
Set the timeout value, in milliseconds, after the first character is received on the
serial line, before data is sent on the network. Valid range is 1 to 30000 milliseconds.
Default is 1000. This setting becomes available when the Timeout mode is selected.
Tunnel Serial
Settings (continued)
Description