Wiring the Profibus Module
4.13.2
Warning
Only a qualified person should make electrical
connections to the system.
Ensure the power is isolated.
Cable glands should meet the requirements of
Gland Selection and Cord Anchorage
Cable insulation and conductor sizes must meet the
requirements of
Cable Conductor Sizes and Insulation
If using armored cable, the armor must not be used as
the main earth connection for the BIOGAS 3000. If
earthing of the armor is required, this must not be taken
from the BIOGAS 3000.
The power supply for the Modbus to Profibus converter
module must not be taken from the BIOGAS 3000. Using
the supply from the BIOGAS 3000 could make the
equipment unsafe resulting in a hazard and invalidate
the ATEX/IECEx certification.
All cables should be crimped with an appropriate
insulated ferrule for the size of the cable being used. In
addition, the cable insulation must be housed
adequately within the protective sheath of the ferrule.
Terminals that are wired should be tightened to a
minimum of 0.6N
m. Failure to tighten to this
requirement could make the equipment unsafe
resulting in a hazard and invalidate the ATEX/IECEx
certification.
The input voltage range to the BIOGAS 3000 Modbus
terminals must not exceed -7 to +12V and the current
must not exceed 83.3mA. Operating outside of this
range will invalidate the ATEX/IECEx certification.