Pool Professional's Meter!

Tracer

TM

 Salt/TDS/Temp 

PockeTester

Code 1749-KIT-01

Содержание Tracer 1749-KIT-01

Страница 1: ...Pool Professional s Meter TracerTM Salt TDS Temp PockeTester Code 1749 KIT 01...

Страница 2: ...Warning This set contains chemicals that may be harmful if misused Read cautions on individual containers carefully Not to be used by children except under adult supervision...

Страница 3: ...ge the TDS Conversion Ratio 7 Changing the Display Temperature Units 8 Data Hold 8 Auto Power Off 8 Low Battery Indicator 8 Testing Getting Started 9 Changing the Measurement Function 9 Measurement 10...

Страница 4: ...Salinity Range 0 to 9 990 ppm Salt fixed ratio of 0 5 TDS Ratio 0 4 to 1 0 adjustable Salinity Ratio 0 5 fixed Conductivity ATC 2 0 per C Temperature Range 32 0 to 149 0 F 0 0 to 65 0 C Temperature R...

Страница 5: ...TRACER PocketTester Code 1749 Calibration Standard 3 000 ppm Code 6005 J Sample Cup w cap Code 1745 1 PARTS ACCESSORIES SAL EC TDS Replacement Electrode Code 1765 Weighted Stand w Sample Cups 5 Code...

Страница 6: ...emperature units recall data 5 ON OFF button 6 Electrode Collar 7 Electrode Note The Electrode cap is not shown TRACER Display 1 Bar graph reading 2 Measurement reading 3 BAT low battery indicator 4 T...

Страница 7: ...y a known conversion ratio factor The meter allows the selection of a conversion ratio factor in the range of 0 4 to 1 0 The selected ratio will vary with application but is typically set between 0 5...

Страница 8: ...10 minutes after the last button was pressed To disable the auto off feature 1 Press the ON OFF button to turn the meter on 2 Wait for the SELF display screen to appear This is the second screen to a...

Страница 9: ...on The meter can be set to measure Conductivity TDS ppm TDS mg L or Salinity ppm To change the mode 1 Turn the TRACER on 2 Press and hold the MODE HOLD button for 2 seconds The display will begin to s...

Страница 10: ...les 6 The meter will autorange to the proper range and the reading will be displayed 7 Rinse the electrode in distilled water Replace the cap Storing Readings 1 After the reading is displayed press an...

Страница 11: ...that the storage recall mode is active 2 The last stored reading taken will be displayed first To advance to the previously stored readings press the MODE HOLD button The location number is displayed...

Страница 12: ...r salinity samples within the range of 1 000 to 9 999 ppm salinity calibrate with a 3 000 ppm calibration standard 1 Fill a sample cup to 20 mL line with a calibration standard 2 Press the ON OFF butt...

Страница 13: ...mory Retrieval On Any Momentary press of CAL RECALL button followed by a momentary press of MODE HOLD button within 4 seconds Scroll Stored Readings On Memory Recall Momentary press of MODE HOLD butto...

Страница 14: ...1 from 0 4 to 1 0 Exit CON TDS Ratio On TDS ppm or mg L Momentary press of CAL RECALL button Change Temperature Units On n a Press and hold CAL RECALL button then momentarily press ON OFF button Relea...

Страница 15: ...move water soluble contaminants from the electrode soak the electrode in deionized water and scrub with a soft brush Battery Replacement 1 Twist off the battery compartment cap 2 Hold the battery hous...

Страница 16: ...ase air bubbles Dirty probe Clean probe Damaged probe Replace probe Contaminated standards Use fresh standards Meter will not turn on Batteries low or dead Replace batteries Battery polarity Replace b...

Страница 17: ...nd proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper ma...