background image

Negative Results

The 2020e and 2020i turbidity meters will display 
negative results if the sample water is less turbid 
than the blank reading.  This is possible in ultra-
low turbidity waters, where the sample turbidity is 
at or near the detection limit of the meter. 

Negative readings can also indicate a problem 
when the meter was blanked.  If the meter is 
accidentally blanked to a 1.0NTU standard, 
subsequent tests may negative, in response to 
the high blank reading.  Correct this problem as 
follows:

1.  Reset the meter to the factory calibration 
settings: with the meter turned OFF, hold down the 
*OK button and press ON.  Release both buttons 
at the same time.  Press *OK to select the default 
settings.  Press *OK again and the meter will turn 
off.  The factory settings will be restored.

2. Follow the instructions on how to properly fill 
the turbidity tubes with the standard solutions (see 
below).

3. Once the tubes are filled with standard, turn 
the turbidity meter on and enter the measurement 
mode.  Scan the blank tube (0.00 NTU).  Then 
leave the blank tube in the meter and press *OK to 
scan the blank as a sample.  It should read 0.00.  
If not, press the UP button and reblank the meter 
again.  Then scan the blank as a sample again 
until it reads 0.00.  This may take at least three 
tries. 

NOTE: a small negative number will be observed 
if the reading is slightly less than the reading 
used as a blank.  This is expected due to minute 
variations between readings.  A reading of -0.02 or 
even -0.04NTU is not uncommon.  Once 0.00 or a 
very small negative number is achieved, the meter 
is ready to be calibrated.  

Filling Turbidity Tubes

It is important to use the proper test tube designed 
for use with the 2020e and 2020i turbidity meters, 
LaMotte code 0475.  Tubes must be clean and free 
from fingerprints, lint, dried spills and significant 
scratches.  Use lint free tissue to wipe your tubes 
clean.

When filling a tube with standard solution, rinse the 
vial with a small amount of the standard solution, 
three times, then fill to the line.  This will assure 
that no dust, debris or dirt will contaminate the 
standard solution that you pour into the vial.  Put 
a dry tube positioning ring on the tube.  The tube 
positioning ring has two tapered notches and one 
squared-off notch.  Place the ring on the tube so 
that the one squared-off notch is aligned with the 
vertical, white indexing line that extends 90° from 
the horizontal fill line on the tube.  This will assure 
that you properly index the tube inside the light 
chamber each time.  Wipe any liquid that may 
have leaked between the positioning ring and tube 
surface.   Cap the tube and apply an Avery label 
marked with the corresponding NTU value.  

Unstable Readings

Unstable readings can occur when excessive 
physical change occurs in the sample at the time 
that the sample is scanned in the meter.  Settling 
and sedimentation can cause this in high turbidity 
waters and even microscopic air bubbles and 
submicron particulates can cause this in low 
turbidity waters.  The ERR4 “Processing Error” 
may be triggered in response to these situations.

To obtain the most accurate and stable readings 
possible, utilize the signal averaging feature 
of the meter.  Turn the meter on and select 
Options, then Averaging.  Choose between 2, 5 
or 10 measurements.  Press OK, then press the 
back button to return to the main menu.  Signal 
averaging of 5 measurements is usually sufficient 
and provides optimal balance of the signal 
averaging benefits, with minimal stress to the 
battery.  

Power Problems

Be sure to use brand name 9V alkaline batteries.  
Heavy Duty, Rayovac and store brand batteries 
do not have enough sustained power to keep 
the meter on and power the lamp.  We suggest 
Duracell Alkaline 9V batteries.  Rechargeable 
batteries can be used; we suggest Duracell or 
Energizer 9V NiMH rechargeable batteries.  If 
using the meter in a lab setting, remove the 
battery altogether and operate exclusively with the 

TC-3000 Meter Troubleshooting Guide

Отзывы: