DC1500-CL 

Chlorine Colorimeter Kit • DPD Tablet

Code  3240

[email protected]

www.

GlobalScientificSupply

.com

GLOBAL SCIENTIFIC SUPPLY

The Laboratory Supply Company!

Содержание DC1500-CL

Страница 1: ...DC1500 CL Chlorine Colorimeter Kit DPD Tablet Code 3240 sales GlobalScientificSupply com www GlobalScientificSupply com GLOBAL SCIENTIFIC SUPPLY The Laboratory Supply Company...

Страница 2: ...Clock 15 Setting Power Save 16 Setting Backlight Time 18 Factory Reset 19 Selecting Language 19 Data Logging 21 Computer Connection 23 Battery AC Operation 23 Maintenance Cleaning 25 Repairs 25 Meter...

Страница 3: ...ecautions 32 Limits of Liability 32 Specifications 33 CE 33 IP67 34 Warranty 34 Register Your Meter 34 Trouble Shooting Guide Trouble Shooting 35 Stray Light 35 sales GlobalScientificSupply com www Gl...

Страница 4: ...to www lamotte com Search the four digit reagent code number listed on the reagent label in the contents list or in the test procedures Omit any letter that follows or precedes the four digit code num...

Страница 5: ...chlorine samples fill sample containers to the top and cap tightly Analyze samples as soon as possible after collection Samples to be analyzed for bromine cannot be preserved or stored INTERFERENCES...

Страница 6: ...o select Testing Menu Testing Menu All Tests Menu Absorbance 12 00 00 001 500 4 Press ENTER to select All Tests Menu All Tests Chlorine TT Chlorine Liq DPD Bromine TT Bromine Liq DPD 12 00 00 001 500...

Страница 7: ...esult as Free Chlorine Chlorine TT 0 99 ppm Scan BLank Scan Sample 12 00 00 001 500 PROCEDURE COMBINED CHLORINE 10 Remove tube from colorimeter Add one Chlorine DPD 3 Instrument Grade Tablet 6197A to...

Страница 8: ...he water sample 12 Subtract free chlorine reading from total chlorine reading to obtain concentration of combined chlorine 13 Press to turn off the colorimeter or press EXIT to exit to a previous menu...

Страница 9: ...select Testing Menu Testing Menu All Tests Menu Absorbance 12 00 00 001 500 4 Press ENTER to select All Tests Menu All Tests Chlorine TT Chlorine Liq DPD Bromine TT Bromine Liq DPD 12 00 00 001 500 5...

Страница 10: ...or press EXIT to exit to a previous menu or make another menu selection WARNING Reagents marked with an are considered to be potential health hazards To view or print a Safety Data Sheet SDS for thes...

Страница 11: ...ding It is not necessary to scan a blank each time the test is performed To use the previous blank reading instead of scanning a new one scroll to Scan Sample and proceed For the most accurate results...

Страница 12: ...ge of the bromine test 1 ppm chlorine 2 25 ppm bromine Follow the procedure below but select the bromine test Chlorine Secondary Standards Code 4140 03 are available to verify the performance of the m...

Страница 13: ...chamber close lid and select Scan Blank 7 Remove tube from colorimeter Add one Chlorine DPD 1 Instrument Grade Tablet 6903A Cap tube and shake for 10 seconds Invert slowly 5 times Solution will turn p...

Страница 14: ...001 500 12 Press ENTER to select Calibrate Two menu choices will be offered Set Calibration and Factory Setting Chlorine TT 1 00 ppm Set Calibration Factory Setting 12 00 00 001 500 13 Press ENTER to...

Страница 15: ...Menu Editing Menu Set Clock Logging Factory Setup Set PWR Save 12 00 00 001 500 4 Press ENTER to select Set Clock The date is displayed as month day year The time is displayed as hours minutes second...

Страница 16: ...is 5 minutes To change the setting 1 Press and briefly hold to turn the meter on The LaMotte logo screen will appear for about 3 seconds and the Main Menu will appear Main Menu Testing Menu Editing Me...

Страница 17: ...to scroll to desired setting Auto Shutoff Disable 5 Minutes 15 Minutes 30 Minutes 12 00 00 001 500 7 Press ENTER to save the selection The screen will display Storing for about 1 second and return to...

Страница 18: ...er The longer the backlight is on the more frequently the battery will have to be charged if the USB Wall Charger is not being used 1 Press and briefly hold to turn the meter on The LaMotte logo scree...

Страница 19: ...00 00 001 500 FACTORY SETUP The Factory Setup menu is used in the manufacturing of the DC1500 Colorimeter This menu is not for use by the operator in the field SELECTING A LANGUAGE There are two langu...

Страница 20: ...to Select Language Editing Menu Factory Setup Set PWR Save Set Backlight Time Select Language 12 00 00 001 500 5 Press to scroll to Select Language Editing Menu English Spanish French Portuguese 12 00...

Страница 21: ...n which it was collected The meter will log the last 500 data points The counter in the center bottom of the display will show how many tests have been logged The display will show 500 when the data l...

Страница 22: ...Menu Set Clock Logging Factory Setup Set PWR Save 12 00 00 001 500 5 Press ENTER to select Logging Logging Display Test Log Enable Logging Disable Logging Erase Log 12 00 00 001 500 6 Press or to scro...

Страница 23: ...r data from the meter to a computer plug the smaller end of the USB cable USB mini B connector into the meter and the larger end of the USB cable USB Type A connector into a USB port on a computer The...

Страница 24: ...ay be degraded A Shutdown Low Batt message on the display will appear for a few seconds before the power is switched off when the battery is too low to operate the unit To extend the battery life Shut...

Страница 25: ...including phone and FAX numbers to the shipping carton This information will enable the service department to make the required repairs more efficiently METER DISPOSAL Waste Electrical and Electronic...

Страница 26: ...ge using a PC The DC1500 may be interfaced with any Windows based computer by using the LaMotte SMARTLink 3 Program GENERAL OPERATING INFORMATION The operation of the DC1500 is controlled by the menu...

Страница 27: ...oves up or down each time an arrow button is pressed Some menus in the DC1500 are looping menus The top and bottom menu choices are connected in a loop Scrolling down past the bottom of the menu will...

Страница 28: ...ms if applicable Footer In the data logging mode the number of the data point is displayed and the total number of data points in the memory will be shown The footer also shows current time and batter...

Страница 29: ...me orientation each time Chambers which have been scratched through excessive use should be discarded and replaced with a new one SAMPLE DILUTION TECHNIQUES If a test result is out of the range of the...

Страница 30: ...nu Absorbance 12 00 00 001 500 3 Press or to scroll to Absorbance Testing Menu All Tests Menu Absorbance 12 00 00 001 500 4 Press ENTER to select Absorbance Absorbance Absorbance 428 Absorbance 525 Ab...

Страница 31: ...s ENTER to scan the sample again The last blank scanned is used by the colorimeter for repeated scans A different blank can be used by pressing or to scroll to Scan Blank and then scanning another bla...

Страница 32: ...ers Some labels include precautionary notices and first aid information Certain reagents are considered potential health hazards and are designated with a in the instruction manual To view or print a...

Страница 33: ...or USB computer connection or lithium ion rechargeable battery Battery Charge Life Approximately 380 tests with backlight on to 1000 tests with backlight off Electrical Ratings Provided on nameplate...

Страница 34: ...rge battery or use USB wall computer charger Shut Down Low Batt Shut Down Battery is too low to operate the unit Charge battery or use USB wall computer charger Overrange Sample is outside of acceptab...

Страница 35: ...Ave Chestertown Maryland 21620 USA 410 778 3100 800 344 3100 www lamotte com 3240 MN 09 09 15 sales GlobalScientificSupply com www GlobalScientificSupply com GLOBAL SCIENTIFIC SUPPLY The Laboratory S...