ANALYSIS
PROCEDURE
1. Fill Water Sample Collecting Bottle (0688) with sample water.
2. Press and hold
until colorimeter turns on.
3. Press
ENTER
to select
Testing Menu
.
4. Scroll to and select
008 Ammonia-N HR
from menu.
5. Rinse tube (0290) with sample water. Fill to the 10 mL line.
Cap and wipe dry.
6. Insert tube. Close lid. Select
Scan Blank
.
7. Remove tube. Add 8 drops of Ammonia Nitrogen Reagent #1.
Cap and mix.
8. Use 1.0 mL pipet (0354) to add 1.0 mL of *Ammonia Nitrogen
Reagent #2.
9. Cap and invert to mix. Wait 5 minutes for full color
development.
10. Insert tube into chamber. Close lid. Select
Scan Sample
.
Record result as
AMMONIA-NITROGEN
.
See manual for calculations.