background image

ANALYSIS

PROCEDURE 

1.  Fill Water Sample Collecting Bottle (0688) with sample water.

2.  Press and hold 

 until colorimeter turns on.

3.  Press 

ENTER

 to select 

Testing Menu

.

4.  Scroll to and select

 008 Ammonia-N HR 

from menu.

5.  Rinse tube (0290) with sample water. Fill to the 10 mL line. 

Cap and wipe dry.

6.  Insert tube. Close lid. Select 

Scan Blank

.

7.  Remove tube. Add 8 drops of Ammonia Nitrogen Reagent #1. 

Cap and mix.

8.  Use 1.0 mL pipet (0354) to add 1.0 mL of *Ammonia Nitrogen 

Reagent #2.

9.  Cap and invert to mix. Wait 5 minutes for full color 

development.

10.  Insert tube into chamber. Close lid. Select 

Scan Sample

Record result as 

AMMONIA-NITROGEN

.

See manual for calculations.

*WARNING: Reagents marked with an * are considered to be potential 

health hazards. To view or print a Safety Data Sheet (SDS) for these 

reagents go to www.lamotte.com. Search for the four digit reagent 

code number listed on the reagent label, in the contents list or in the 

test procedures. Omit any letter that follows or precedes the four digit 

code number. For example, if the code is 4450WT-H, search 4450. To 

obtain a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from 

Chem-Tel (US, 1-800-255-3924) (International, call collect, 813-

248-0585).

DC1500  

AMMONIA

 

Quick Guide