Instructions, Silicon Diode Temperature Sensor 
Installation, Model DT-470, 471, 670, and 671 CY Package 

 

 

 

Lake Shore Cryotronics, 575 McCorkle Blvd. Westerville, OH 43082 

www.lakeshore.com

  

Phone (614) 891-2243 

– Fax (614) 891-1362 – 

[email protected]

 

 

[email protected]

  

CAUTION: 

This temperature sensor is sensitive to electrostatic discharge (ESD). Use ESD 

precautionary procedures when handling, or making mechanical or electrical connections to this 
device in order to avoid performance degradation or loss of functionality. 

 

 
There are three aspects of using a cryogenic temperature sensor which are critical to its optimum performance. The first 
involves the proper mounting of the sensor package; the second relates the proper joining of sensor lead wires and 
connecting wires. The final concern is the thermal anchoring of the lead wires. Although the sequence in which these areas 
should be addressed is not set in stone, all elements covered under 
each aspect should be adhered to for maximum operating capabilities 
of the sensor. 
 

SENSOR MOUNTING 

The DT-470/471/670/671 CY package combines a standard SD sensor 
with a gold-plated copper mounting bobbin to form the CY package. 
The mounting bobbin of these packages contains a 3.1 mm (0.122 in) 
diameter centered hole designed for mounting with a #4-40 or M3 
screw. 
 
1.  A threaded hole in your mounting surface is necessary for 

mounting the sensor package. The hole in the sensor package will 
accommodate a #4-40 or M3 screw. A brass screw is recommended due to the thermal contractions/expansions of the 
final assembly. 
 

2.  The threaded hole and surrounding surface should be cleaned with a solvent such as acetone followed by an isopropyl 

alcohol rinse. Allow time for the solvents to evaporate before sensor mounting. 
 

3.  Apply a small amount of Apiezon

®

 N grease to the threads of the screw. To ensure good thermal contact between the 

sensor and mounting surface use an indium washer/preform or a thin layer of Apiezon

®

 N grease between the mounting 

surface and the sensor package. 

 

NOTE:

 An overabundance of grease will increase the thermal barrier. Keep the thickness to 0.05 mm or less. 

 

4.  Insert screw through sensor mounting bobbin and tighten screw firmly enough to hold sensor in place. Avoid over 

tightening (torque of 3 to 5 in-oz [0.02 to 0.035 N-M] should be sufficient). 
 

NOTE:

 This package is designed for use up to 400 K (127 °C); subjecting sensor to 

temperatures in excess of this will cause a shift in the sensor values.  
 
 

LEAD ATTACHMENT 

The SD sensor has been attached to the mounting bobbin and encapsulated in Stycast

®

 epoxy. 

The 914 mm (36 in) sensor leads are 30 AWG, Teflon

®

 coated wire which are thermally 

anchored to the bobbin. Standard lead configuration for the CY is a 2-lead device  
[Green = Cathode (

–), Yellow = Anode (+)]. 

 
914 mm (36 in) lead wire is attached during the production process. If additional connection wire is required, use the following 
instructions. 
 
1.  Prepare the sensor leads with a RMA (rosin mildly active) soldering flux. Use a low wattage soldering iron and tin leads 

with a minimal amount of 60% Pb 40% Sn solder. Clean off residual flux with a rosin residue remover. 
 

2.  Strip the insulation from the connecting wires by delicately scraping with a razor blade, fine sand paper, or steel wool. 

Follow the same procedure in step 1 for preparing connecting wires.  

NOTE:

 This package is designed for use up to 400 K (127 °C), subjecting sensor to temperature in excess of this will 

cause a shift in the sensor values.  
 
 

 