3.6.4 Scanner Output
Connector Definition
Two 6-pin (240°) circular DIN 45322 receptacles are used for the resistance measure-
ment output from the scanner. These are identical to the connectors used on the
measurement input of the Model 372. The cables included with the scanner are ade-
quate for most installations. In the event that you must make a cable for a special
application, connector pin definition on the scanner is identical to the input connec-
tor definition given in section 3.5.1. Guard pins are not connected inside the scanner
but guards can be carried through the cable making them reversible. Connectors
mate with 6-pin circular DIN plug (Lake Shore P/N G-106-233).
3.6.5 Scanner Output
Cabling
A scanner to instrument cable (Lake Shore P/N G-112-374) is included with the 3726
and 3708 scanners. The figure below illustrates the cable wiring. This configuration is
recommended when making cables for the scanner. Shielded twisted pair cables, as
described in section 3.5.3, can be used instead of dual triaxial cables for less demand-
ing applications. Cable shields should always be tied to connector shield pins on both
ends of this cable.
FIGURE 3-9
Scanner input connector
Pin
Symbol
Pin
Symbol
1
Shield
14
Shield
2
Shield
15
1 I–
3
1 I+
16
1 V–
4
1 V+
17
Shield
5
Shield
18
2 I–
6
2 I+
19
2 V–
7
2 V+
20
Shield
8
Shield
21
3 I–
9
3 I+
22
3 V–
10
3 V+
23
Shield
11
Shield
24
4 I–
12
4 I+
25
4V–
13
4 V+
Body
Shield
Guards are not carried through the scanner. Shields are connected to isolated
measurement common.
TABLE 3-4
Scanner input connector details
FIGURE 3-10
Scanner to instrument measurement cable